6. Over pressure valve
Press and hold
Options
, then press
On/Off
.
When the
Enter ID
display appears, enter
3031
.
When the
Enter Password
display appears, enter
7867
.
Press
Options
until
Pressure Calibrate
appears in display.
Lower the end of the tubing to the floor (approximately 30 inches
below the pump).
Program in 100 mL/h (99.0 mL/h on
3100/505
) and run the pump until
fluid is flowing from end of tubing. When the motor position value
(left side of display) is 1650 ± 50, press
Hold
.
Watch the end of the tubing for 30 seconds to verify that no more than
1 drop is expelled.
7. Valve integrity
Connect the regulated compressed air source to the lower tubing.
While maintaining the motor position value from step 6 above (1650 ±
50, top valve occluded) apply 12 psi back pressure to the tubing.
Verify that no fluid/air movement occurs in the tubing (allow the
tubing expansion created by the back pressure to stabilize).
Run the pump until the motor position value is 700 ± 50 (middle valve
occluded).
Press
Hold
and verify that no fluid/air movement occurs in the tubing
(allow the tubing expansion created by the back pressure to stabilize).
8. Pressure plate position micro switch test
Reduce the back pressure and remove the cassette from the pump.
Hang a 2.5 long 0.015 feeler gauge onto the upper left cassette
loading pin (gauge must hang within the cassette housing recession).
Close the door and verify that the cassette position number (right-
hand side of the display) changes from 0 to 1.
Open the door and remove the 0.015 feeler gauge. Hang a 2.5 long
0.012 feeler gauge in same manner as the 0.15 gauge.
Close the door and verify that the cassette position number (right-
hand side of the display) remains at 0.
Remove the feeler gauge from pump.
If the pump fails this test then carry out the adjustment procedure on
page 7-32.
9. Audio alarm
Press
Options
and advance to the alarm test display. Press
1
to test
the
non-insistent
audio
alarm. Press
2
to test the
insistent
audio
alarm, and press
3
to test the
Back Up
alarm. Turn the pump off.
Содержание 3000/500
Страница 1: ...0473 Technical Service Manual Model 3000 500 and Micro 3100 505 Volumetric Infusion Pump...
Страница 2: ......
Страница 14: ...I n t r o d u c t i o n Chapter 1...
Страница 18: ...S p e c i f i c a t i o n Chapter 2...
Страница 29: ...O v e r v i e w o f P u m p O p e r a t i o n I n i t i a l C h e c k s Chapter 3...
Страница 47: ...V o l u m e t r i c P u m p M e n u s Chapter 4...
Страница 79: ...M e c h a n i c a l S y s t e m s Chapter 5...
Страница 95: ...E l e c t r o n i c C i r c u i t s Chapter 6...
Страница 96: ...M a i n t e n a n c e P r o c e d u r e s T r o u b l e s h o o t i n g Chapter 7...
Страница 129: ...D i s a s s e m b l y R e a s s e m b l y C a l i b r a t i o n I n s t r u c t i o n s Chapter 8...
Страница 151: ...I l l u s t r a t e d P a r t s L i s t Chapter 9...
Страница 178: ...E l e c t r i c a l D i a g r a m s Chapter 10...
Страница 189: ...W a v e f o r m D i a g r a m s Chapter 11...
Страница 198: ...R e m o t e M o n i t o r i n g C o n t r o l o f t h e p u m p Chapter 12...
Страница 225: ...I n d e x...