4 — 1 7
G r a s e b y M e d i c a l L t d .
4
3 0 0 0 / 3 1 0 0 a n d 5 0 0 / 5 0 5 S e r v i c e M a n u a l
I s s u e A ( A p r i l 2 0 0 2 )
V o l u m e t r i c I n f u s i o n P u m p s
Chart A Pump Main Modes
Once you have converted the decimal value in field AAA into binary
format, use this chart to find out the pumps state modes. The
example of Snapshot-b shows field AAAs value as 072 (decimal),
which converts into 01001000.
Bit 7
Bit 6
Bit 5
Bit 4
Bit 3
Bit 2
Bit 1
Bit 0
0 =
Not used
Unit not
Unit not
Unit not
Unit not
Unit not
Unit is in Unit not
in dose-
in quick
in dead
in low
in KVO
primary
running
rate calc
rate change battery
battery
condition
condition
1=
Not used
Unit is in Unit is in Unit is in Unit is in Unit is in Unit is in Unit is
dose-rate
quick rate dead
low battery KVO
secondary running
calc.
change
battery
condition
condition
Chart B Pump State Modes
This chart shows the state modes that can be decoded from field BBB
on Snapshot-b when the value is converted to binary. The example of
Snapshot-b shows a value of 004 (decimal) which converts into
00000100.
Bit 7
Bit 6
Bit 5
Bit 4
Bit 3
Bit 2
Bit 1
Bit 0
0 = Primary/
Secondary Secondary Primary
Primary
Standard
Pump not Pump not
secondary
VTBI does rate does
VTBI does rate does
message
in rate
in rate
rate/volume not have
not have
not have
not have
not in
taper end
taper
not decimal decimal
decimal
decimal
decimal
message
early
since total
place
place
place
place
display
volume last
cleared
1= Primary/
Secondary Secondary Primary
Primary
Standard
Pump is in Pump is in
Secondary
VTBI has
rate has
VTBI has
rate has
message is rate taper rate taper
rate/volume decimal
decimal
decimal
decimal
in message end early
was decimal place
place
place
place
display
since total
volume last
cleared
Chart C Pump Physical Modes
Convert the value in CCC to binary then use this table to decode the
bits. In the example screen, the decimal value is 029, or binary
00011101.
Bit 7
Bit 6
Bit 5
Bit 4
Bit 3
Bit 2
Bit 1
Bit 0
0 =
Not used
Run not
Not used
Cassette not Not used Safety clip
Door not Pump not
pendingin position
not in place open
turned on
1 =
Not used
Run
Not used
Cassette in
Not used Safety clip
Door open Pump
pendingposition
in place
turned on
Содержание 3000/500
Страница 1: ...0473 Technical Service Manual Model 3000 500 and Micro 3100 505 Volumetric Infusion Pump...
Страница 2: ......
Страница 14: ...I n t r o d u c t i o n Chapter 1...
Страница 18: ...S p e c i f i c a t i o n Chapter 2...
Страница 29: ...O v e r v i e w o f P u m p O p e r a t i o n I n i t i a l C h e c k s Chapter 3...
Страница 47: ...V o l u m e t r i c P u m p M e n u s Chapter 4...
Страница 79: ...M e c h a n i c a l S y s t e m s Chapter 5...
Страница 95: ...E l e c t r o n i c C i r c u i t s Chapter 6...
Страница 96: ...M a i n t e n a n c e P r o c e d u r e s T r o u b l e s h o o t i n g Chapter 7...
Страница 129: ...D i s a s s e m b l y R e a s s e m b l y C a l i b r a t i o n I n s t r u c t i o n s Chapter 8...
Страница 151: ...I l l u s t r a t e d P a r t s L i s t Chapter 9...
Страница 178: ...E l e c t r i c a l D i a g r a m s Chapter 10...
Страница 189: ...W a v e f o r m D i a g r a m s Chapter 11...
Страница 198: ...R e m o t e M o n i t o r i n g C o n t r o l o f t h e p u m p Chapter 12...
Страница 225: ...I n d e x...