Advanjet HV-2000C Jet Controller Setup and Operation
3A5856A
7.6
Output Commands
OD
Output Internal Drop Counter
OD;
The host can send the
OD;
command to determine the number of drops that have been
completed since the last
SD;
command (reset counter command). A return string
consisting of the total number of drops will be returned.
OE
Output Error
OE;
The host can send the
OE;
command to verify that it is communicating properly with the
controller. A return string of
0;
indicates that the command was received without error
and
1;
indicates there was an error in the previous command. Once the
OS
; command
return string is returned, the error state in the controller is cleared.
OS
Output Jet Status
OS;
The host can send the
OS;
command to verify that it is communicating properly with the
controller. A return string of
1;
indicates that the Jet valve is closed and
0;
indicates the
valve is opened.
OT
Output Temperature
OT;
The host can send this
OT;
command to get the present temperature
of the Jet heater.
A return string of
50.5;
indicates the temperature is at 50.5 degree Celsius.
OV
Output Version
OV;
The host can send this
OV;
command to identify the controller. A return string of
ADV-HV2000P_xxxx
indicates that the host has found the controller.
Note:
The controller might not be able to respond to the output commands when
it is busy dispensing a long series of drops. In this case, the host should try again
after a short wait.