Troubleshooting
334852D
7
Troubleshooting
NOTE:
•
Relieve pressure.
See pressure relief procedure
included in your dispense valve instruction manual.
•
Before you check or repair the meter, be sure the
pump and all other valves and controls are operat-
ing properly.
Problem
Cause
Solution
Digital display is dim.
Battery in electronic control is worn
out.
Replace battery. See page 8.
Electronic control is malfunctioning.
Replace the electronic control.
Order Part No. 24V554.
Digital display does not activate.
Battery in electronic control is worn
out.
Replace battery. See page 8.
Electronic control is malfunctioning.
Replace the electronic control.
Order Part No. 24V554.
There is no fluid flow.
Strainer, if used is clogged.
Remove and clean strainer.