16
M0303
EN
K 4.2.1
IN-FIELD CALIBRATION PROCEDURE
ACTION
DISPLAY
1
NONE
NEXT in Standby
2
CAL
CAL
CAL
LONG CAL key keying
The NEXT enters calibration mode, shows <<CAL>> and displays
the calibration factor in use instead of partial. The words “Fact” and
“USER” indicate which of the two factors (factory or user) is currently
in use.
Important: This factor is that which the instrument also uses for fiel
calibration measurement operations
3
RESET
RESET
RESET
LONG RESET key keying
The NEXT shows “CAL” and the partial at zero. The NEXT is ready to
perform in-field calib ation.
4
DISPENSING INTO SAMPLE CONTAINER
Without pressing any key, start dispensing into the sample container
Dispensing can be interrupted and started again at will. Continue
dispensing until the level of the fluid in the sample ontainer has
reached the graduated area. There is no need to reach a preset
quantity.
Indicated value Real value
5
RESET
SHORT RESET key keying
The NEXT is informed that the calibration dispensing operation is
finishe .
Make sure dispensing is correctly finished before performing this operation.
To calibrate the NEXT, the value indicated by the partial totaliser (example
9.800) must be forced to the real value marked on the graduated sample
container. In the bottom left part of the display an arrow appears (upwards
and downwards), that shows the direction (increase or decrease) of the value
change displayed when the following operations 6 or 7 are performed.
6
RESET
SHORT RESET key keying
The arrow changes direction. The operation can be repeated to
alternate the direction of the arrow.
7
CAL
CAL
CAL
CAL
SHORT/LONG CAL key keying
The indicated value changes in the direction indicated by the arrow
- one unit for every short CAL key keying
- continually if the CAL key is kept pressed. The speed increase rises
by keeping the key pressed. If the desired value is exceeded, repeat
the operations from point (6).
8
RESET
RESET
RESET
NOTICE
LONG RESET key keying
The NEXT is informed that the calibration procedure is finishe .
Before performing this operation, make sure the INDICATED value is
the same as the REAL value.
.
Indicated value Real value
The NEXT calculates the new USER K FACTOR ; this calculation could
require a few seconds, depending on the correction to be madeù
If this operation is performed after action (5), without
changing the indicated value, the USER K FACTOR would
be the same as the FACTORY K FACTOR, thus it is ignored.
Содержание 127650
Страница 22: ...22 M0303 EN Q DIMENSIONS 190 160 79 433 190 19 Ø 20 Ø 3 4 HOSE ...
Страница 25: ...Cette page est intentionnellement laissée en blanc ...
Страница 52: ...Esta página se ha dejado en blanco intencionadamente ...
Страница 76: ...76 M0302 ES Q DIMENSIONES TOTALES 190 160 79 433 190 19 Ø 20 Ø 3 4 HOSE ...
Страница 78: ...Esta página se ha dejado en blanco intencionadamente ...
Страница 79: ...Esta página se ha dejado en blanco intencionadamente ...