M0303
15
EN
K4.1
DISPLAY OF CURRENT CALIBRATION
FACTOR AND RESTORING FACTORY FACTOR.
CAL
CAL
CAL
By pressing the CAL key while the meter is in
Standby, the display page appears showing the
current calibration factor used. If no calibration
has ever been performed, or the factory setting
has been restored after previous calibrations,
the following display page will appear: The
word “Fact” abbreviation for “factory” shows
that the factory calibration factor is being used
If, on the other hand, calibrations have been
made by the user, the display page will appear
showing the currently used calibration factor ( in
our example 0,998) . The word “user” indicates a
calibration factor set by the user is being used..
CAL
The fl w chart alongside shows
the switchover logic from one
display page to another
In this condition, the Reset key
permits switching from User fac-
tor to Factory factor.
To confi m the choice of cali-
bration factor, quickly press
CAL while “User” or “Fact” are
displayed.
After the restart cycle, the K24
uses the calibration factor that
has just been confi med
CAL
CAL
CAL
CAL
RESET
RESET
TIME OUT
NOTE
When the Factory Factor is confirmed, the old User factor is
deleted from the memory
K4.2
IN FIELD CALIBRATION
FOREWORD
This procedure calls for the fluid to be dispensed into a graduated sample con-
tainer in real operating conditions ( fl w rate, viscosity, etc.) requiring maximum
precision.
NOTE
For correct meter calibration, it is most important to:
1
When the Factory Factor is confi med, the old User factor is deleted from the
memory
2
use a precise Sample Container with a capacity of not less than 5 litres, featuring
an accurate graduated indicator.
3
ensure calibration dispensing is done at a constant fl w rate equivalent to that of
normal use, until the container is full;
4
Not reduce the fl w rate to reach the graduated area of the container during
the final dispensing stage (the correct method during the final stages of sample
container filling onsists in making short top-ups at normal operation fl w rate) ;
5
after dispensing, wait a few minutes to make sure any air bubbles are eliminated
from the sample container; only read the Real value at the end of this stage, dur-
ing which the level in the container could drop.
6
Carefully follow the procedure indicated below.
Содержание 127650
Страница 22: ...22 M0303 EN Q DIMENSIONS 190 160 79 433 190 19 Ø 20 Ø 3 4 HOSE ...
Страница 25: ...Cette page est intentionnellement laissée en blanc ...
Страница 52: ...Esta página se ha dejado en blanco intencionadamente ...
Страница 76: ...76 M0302 ES Q DIMENSIONES TOTALES 190 160 79 433 190 19 Ø 20 Ø 3 4 HOSE ...
Страница 78: ...Esta página se ha dejado en blanco intencionadamente ...
Страница 79: ...Esta página se ha dejado en blanco intencionadamente ...