78
AED mode
• Energy:
Fixed energy at 200 J.
• Protocol:
ERC 2010 CPR Guidelines with voice and text prompts.
• Shockable rhythms:
VF with amplitude >0.15 mV and VT with rhythm >150 bpm.
• Sensitivit
y:
Shockable rhythm VF > 95%.
Shockable rhythm VT > 75%.
• Specificity:
Normal sinusoidal rhythm > 99%.
Asystole and other non-shockable rhythms > 95%.
SpO
2
(optional)
•
SpO
2
range:
0
–
100%.
•
HR range:
30
–
250 bpm.
•
Accuracy:
70
–
100 % ~ 2% for adults with finger clip sensor.
•
Alarm:
Adjustable min 50 %.