On the bottom side of the screen the function keys (F1…F5) are the commands for the
NIBP:
START/STOP
–
F1
: to start or abort the NIBP measurement in manual or automatic
mode.
–
F2
: to print the last NIBP measurement in manual mode or paper
feed. If the last measurement was already printed it will only
perform paper feed. In automatic operation mode will perform
only paper feed as the last measurement is automatically printed.
CLEAR
–
F3
:
to clear the last measurement values (SYS,DIA,MEAN).
SET
–
F4
:
to enter the NIBP set up menu.
EXIT
–
F5
:
to exit the NIBP and return to the defibrillator functions.
NIBP SET-UP MENU
Pressing the
SET
–
F4
key the NIBP set-up menu is displayed.
To modify the values press the selection jog to choose the field then rotate the selection
jog to change the value assigned to the field.
To exit the set-up menu press the
SET
–
F4
key.
PATIENT:
select the patient type: ADULT, PEDIATRIC or NEONATAL.
MODE:
select the measurement mode: MANUAL or AUTOMATIC.
AUTO INTERVAL: select the automatic measurement interval: 1,2,3,4,5,10,15,30,60
or 90min.
PRINT MODE:
select manual or automatic hardcopy only for MANUAL measurement
mode. In automatic measurement mode the printing is always
automatic.