![Gima Rescue LIFE Скачать руководство пользователя страница 40](http://html1.mh-extra.com/html/gima/rescue-life/rescue-life_user-manual_2228086040.webp)
Distributed by
Rescue Life - User Manual Rev. 3.1
34
DEFIBRILLATOR PROCEDURE IN MANUAL MODE
1.
Switch on the device pressing the
ON/OFF
button. Connect and lock the paddles
plug to start the operation. The ECG signal will be displayed and by default the
energy is set to
150 J.
2.
Place the paddles (set in adult or in pediatric version) on the patient chest and
analyze the ECG trace in order to decide if defibrillation is necessary.
3.
Select the energy level required using the Speed Dial.
4.
If you are using the standard paddles press both push buttons on the handles to
start charging.
5.
If you are using disposable pads press the
CHARGE
button on the device panel to
start charging.
6.
On the screen the charge status bar indicates that the charging procedure is on; at
the same time the ascending sound will start.
7.
When the charge ends the red light on the
SHOCK
button will turn on indicating
that
RESCUE LIFE
is ready for defibrillation.
8.
To release the defibrillation shock, press both push buttons on the standard
paddles. If using disposable pads press the SHOCK key in the front panel
to release the defibrillation shock.
9.
The shock has to be released within 25 sec from the charge completed; after 25 sec
the RESCUE LIFE will discharge internally.
10.
If the defibrillation is not required, press the
DISARM (F1)
key to discharge
internally.
11.
If the SYNC function is on, SYNCHRONIZED CARDIOVERSION can be done.
DO
NOT
TOUCH
THE
PATIENT
DURING
DEFIBRILLATION.
W
HEN THE MESSAGE
“
ATTACH PADS
”
IS DISPLAYED ON THE SCREEN AND THE CHARGE IS STARTED
,
THE
DEVICE ASSUMES A STANDARD IMPEDANCE OF
50
OHM
.
I
F THE MESSAGE
“
ATTACH PADS
”
PERSISTS WHEN
THE SHOCK HAS TO BE DELIVERED
,
THE DEVICE WILL DISCHARGE INTERNALLY
T
HE MESSAGE
“
ATTACH PADS
”
CAN BE DISPLAYED ALSO IN PRESENCE OF A NO SUFFICIENT ELECTRIC
CONTACT BETWEEN THE PADS AND THE PATIENT SKIN
;
IN THIS CASE ADD CONDUCTIVE GEL AND PRESS
STRONGLY THE DEFIBRILLATION PADS ON THE PATIENT SKIN
.
D
URING
SYNC
MODE
,
THE SHOCK WILL NOT
BE RELEASED IF THE
ECG
TRACE IS NOT STABLE AND THE
QRS
COMPLEX IS NOT VALID
.
W
HEN USING
SYNC MODE
MAKE SURE THAT THE
ECG
TRACE HAS A STABLE BASE LINE AND THE HEART
RATE IS STABLE
.
D
EFIBRILLATING IN
SYNC MODE
WITH A DISTURBED
ECG
SIGNAL IS DANGEROUS
BECAUSE THE MACHINE WILL NOT BE ABLE TO IDENTIFY CORRECTLY THE
‘R’
PEAK TO SYNCHRONIZE TO
.
T
HE MACHINE CAN DELIVER A
SYNC
SHOCK USING AS INPUT THE
ECG
CABLE OR THE PADS
,
BUT IT IS
RECOMMENDED TO USE THE PADS INPUT FOR THE BEST RESULT
.