Distributed by
Rescue Life - User Manual Rev. 3.1
26
The HR BEEP is not stored and when the machine is switched on, will be active (heart
rate beep on). For patient safety reason, it can be set to off only for the actual working
session.
DEVICE AND PATIENT PREPARATION
PRODUCT CHECK
Check carefully the content of the packing for any damage that might have been occured
during shipping.
Check carefully all the accessories to ensure that the unit comes with the complete
accessories necessary for a proper use of the device.
HOW TO TEST THE DEFIBRILLATOR
It is possible to charge without attaching the pads to the patient and discharge internally
from the standard paddles only for the defibrillator testing and only selecting 1 joule.
If the defibrillator is charged using this mode the standard impedance of 50 ohm is
assumed.
ELPRO S.r.l. recommended to make one functional check, preventive
maintenance and electrical safety test at least once a year.
W
HEN THE
RESCUE
LIFE
IS CHARGED WITH THE PADDLES NOT ATTACHED TO THE PATIENT AND THE
ENERGY IS SET TO A VALUE HIGHER THAN
150
J
THE CHARGED ENERGY WILL BE LIMITED TO
150
J.
I
N THIS CASE THE MESSAGE
“ENERGY
LIMIT”
WILL BE DISPLAYED
.
T
O OBTAIN THE BEST SHOCK RESULT IT IS STRONGLY RECOMMENDED TO CHARGE THE ENERGY WITH THE
PADDLES ATTACHED TO THE PATIENT
.
N
EVER CHARGE THE DEFIBRILLATOR WITH THE PADDLES IN CONTACT BETWEEN THEM
.
M
AKE SURE THAT THE CONTACT GEL IS SPREAD ONLY ON THE PADDLES AND NOT ALL OVER PATIENT CHEST
.
I
F THIS INDICATION IS NOT
FOLLOWED BURNS TO PATIENT CHEST AND DEFIBRILLATOR FAULTS MAY OCCUR
.
PROVIDE
AN
INTERVAL
OF
AT
LEAST
60
SECONDS
BETWEEN
THE
CYCLE
OF
CHARGE
/
DISCHARGE
WHEN
THE
ELECTRODES
ARE
NOT
CONNECTED
TO
THE
PATIENT.