Distributed by
Rescue Life - User Manual Rev. 3.1
33
DEFIBRILLATION MODE
With RESCUE LIFE users can choose between three operating modes: manual, advisory
and semiautomatic.
In MANUAL mode users have to analyze heart rhythm and decide time and energy value
to use for the defibrillation while in other modes RESCUE LIFE automatically analyze
heart rhythm.
In ADVISORY mode (ADV) RESCUE LIFE only analyzes and suggest if patient needs
defibrillation, after that users have to choose energy value, charge defibrillator and
deliver the shock.
In SEMIAUTOMATIC mode (AED) RESCUE LIFE analyzes heart rhythm and helps
operators with visible and audible prompts. If patient needs a defibrillation RESCUE LIFE
automatically charge with energy value of 200 Joule.
Operators only should to deliver shock and perform CPR when indicated.