Rescue Life - User Manual Rev. 3.1
47
NIBP (NON-INVASIVE BLOOD PRESSURE)
INTRODUCTION
The
RESCUE LIFE NIBP
option is based on the
Advantage
A+
module platform
available in the series of oscillometric OEM NIBP technologies from
SunTech Medical
®
.
The
Advantage
series of OEM NIBP technologies provides the simplicity of the
oscillometric technique of acquiring blood pressure with the most reliable, flexible and
clinically accurate modules in the industry.
Advantage
OEM NIBP technologies have been integrated in many different medical
devices throughout the world including general multi-parameter patient monitors,
cardiac output monitors, dialysis machines, defibrillators, 24-hour ABPM devices,
anesthetic delivery devices as well as several niche market devices.
The
Advantage
A+ module provides the highest NIBP performance in the smallest
complete package available. SunTech Medical designed the
Advantage
A+ to meet the
needs of the most challenging clinical application with internal automatic modes, low
voltage communication protocols, the lowest power consumption in the industry and
readily equipped to integrate the
Advantage
RMT
technology option
for the highest level
of motion tolerant performance.
OPERATIONAL OVERVIEW
For pediatric and adult patient populations, blood pressure measurements made with
the
Advantage
series of OEM NIBP technologies are equivalent to those obtained by
trained observers using the cuff/stethoscope auscultatory method within the limits
prescribed by ANSI/AAMI SP10: 2002 (mean error difference of ± 5 mmHg or less,
standard deviation of 8 mmHg or less) as well as EN1060-4:2004.
For neonatal patient populations, blood pressure measurements made with the
Advantage
series of OEM NIBP technologies are equivalent to those obtained by intra-
arterial blood pressure devices within the limits prescribed by ANSI/AAMI SP10: 2002
(mean error difference of ± 5 mmHg or less, standard deviation of 8 mmHg or less) as
well as EN1060-4:2004.