Immunity test
Test level to
IEC 60601-1-2
Conformity level
Electromagnetic environment -
directives
Conducted RF
IEC 61000-4-6
Radiated RF
IEC 61000-4-3
3 Veff
150 kHz to 80 MHz
3 V/m
80 MHz to 2.7 GHz
3 Veff
3 V/m
Portable and mobile RF communications
equipment should be used no closer to any
part of the Products, included cables, than
the recommended separation distance
calculated from the equation applicable to
the frequency of the transmitter.
Recommended separation distance:
d = 1,2√P
150 KHz to 80 MHz
d = 1,2√P
80 MHz to 800 MHz
d = 2,3√P
800 MHz to 2.7 GHz
where
P
is the maximum output power
rating of the transmitter in watts (W),
according to the transmitter manufacture
and
d
is the recommended separation
distance in meters (m).
Field strengths from fixed transmitters, as
determined by an electromagnetic site
survey, should be less than the compliance
leave in each frequency range.
Interference may occur
in the vicinity of equipment marked with
the following symbol:
NOTE 1: At 80 MHz and 800 MHz, the higher frequency range applies.
NOTE 2: These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and
reflection from structures, objects and people.
Содержание GIMAled81
Страница 37: ...MANUALE D USO E MANUTENZIONE IT MU_027 Rev 1 29 04 2020 Pag 13 di 33 Modello a piantana...
Страница 57: ...MANUALE D USO E MANUTENZIONE IT MU_027 Rev 1 29 04 2020 Pag 33 di 33 Note...
Страница 94: ...OPERATION AND MAINTENANCE MANUAL EN MU_027 Rev 1 29 04 2020 Page 13 of 33 Mobile model...
Страница 114: ...OPERATION AND MAINTENANCE MANUAL EN MU_027 Rev 1 29 04 2020 Page 33 of 33 Notes...