72
ENGLISH
12.3 Screen Display with ECG Waveforms only
Press the DISP key to shift screen to 7 ECG Waveform Screen when setting Disp2 as “7 ECG” in System Setup screen. In
this screen, the operator can view 7 traces of ECG waveforms simultaneously for lead I, II, III, AVR, AVL, AVF and V, as
shown in Figure 2.9.
Figure 12.3 7 traces of the ECG waveforms display only
Operation Instructions:
ECG lead: No action.
Alarm silence: press it to enable or disable the alarm silence function.
Freeze: press it to freeze all 7 ECG waveforms. Double press this key within 2 seconds to lock or unlock the
operation of all other buttons (except power switch) on the front panel.
NIBP: press it to start or stop NIBP measure.
DISP: press it to shift the display to the Main Screen.
Navigation Knob:
rotate the knob to adjust the gain for all 7 trace of ECG waveforms. The ECG gain includes 6 options:
“Auto”, “X1/4” “X1/2”, “X1”, “X2”, “X4”.
Notes:
1. For the monitor without ECG function, the display view of 7 traces of ECG waveforms will not be available.
2. When “3” lead wires are chosen in “Cable” setting of ECG Setup, the display view of 7 traces of ECG waveforms will
not be available.
3. When “3” lead wires are chosen in “Cable” setting of ECG Setup, the screen changes to the below figure: