Figure 4.20 MC Analysis Screen
Rotate the Navigation knob to move the cursor to the option which needs to be calculated, press the knob and rotate it
to obtain calculating value. When the calculating value is selected, the calculated value will be displayed in corresponding
position. Each calculating option has limit range, if the result exceeds range, it will display “…”.
On MC analysis screen, other menu options can not enter value unless entering patient’s weight and
medicine name again, in default status it is no effective. The values in system is a group of stochastic initial
values, the operator should not consider it as calculating standard, please according to doctor’s device enter
a group values which are suitable for patient.
The unit of every medicine is settled unit or unit series. The operator must select the appropriate unit
according to doctor’s device. In a unit series, unit carry performs automatic adjustment along with the
current entering value. When exceeding the range of this unit expression, the system will display “…”.
When the operator finishes one option entering, the system will give visible indication in menu to remind
operator to check the correctness of entering value.
For every new entering value, please perform confirming. The operator should take it seriously, only the
entering is correct, the calculating result is believable and reliable.
Select Medicine Type:
Move the cursor to “Medicine”, rotate Navigation knob to perform selection. Ten options:
AMINOPHYLLINE, DOBUTAMINE, DOPAMINE, EPINEPHRINE, HEPARIN, ISUPREL, LIDOCAINE, NIPRIDE, NITROGLYCERIN,
and PITOCIN. The default medicine is AMINOPHYLLINE.
Weight:
when entering into medicine calculating window, the operator should enter patient’s weight; the weight is used
for MC calculation only; weight: 0.5Kg to 300Kg selectable; step: 0.5Kg; default: 70 Kg for adult; 20Kg for Pediatric.
Medicine calculation function just supplies a medicine calculator function. The values in table can have no
relation with the monitored patient, so the weight in this menu and the weight in system are two different
values. When update a patient in system operation, the value in this menu will not be affected.