clean wipe or simply air-dry.
This monitor can be disinfected, please clean the monitor firstly.
Do not let the liquid cleanser flow into the connector jack of the monitor to avoid damage.
Clean the exterior of the connector only.
Dilute the cleanser.
Do not use scrub materials.
Do not let any liquid flow into the shell or any parts of the monitor.
Do not let the cleanser and disinfectant stay on its surface.
Do not perform high pressure sterilization to the monitor.
Do not put any parts of the monitor or its accessories in the liquid.
Do not pour the disinfectant on its surface while disinfectant.
If the monitor is accidentally wet, it should be thoroughly dried before use. The rear cover can be removed by
qualified service technician to verify absence of water.
Never use this machine in an environment with inflammable gas.
Avoid being hit by lightning. The power cable should be plugged into an outlet with grounding wire. Do not use
an outlet with poor condition. If possible, use power supply system with regulator.
It must be used in a clean environment protected against shock. Keep it away from corrosive substances,
explosive substances, high temperature and dampness.
If it is installed in a cabinet, make sure the installation allows for good ventilation, and easy maintenance,
observation and operation.
14.2 Disinfecting the Device and Accessories
Disinfection may cause damage to the device and is therefore not recommended for this monitor unless otherwise
indicated in your hospital’s servicing schedule. Cleaning device before disinfecting is recommended.
The recommended disinfects include:
75% Ethanol
70% Isopropanol
Do not use damaged accessories.
Accessories can not be entirely immerged into water, liquor or cleanser.
Do not use radiation, steam or EO to disinfect accessories.
Do wipe off the remained Ethanol or ispropanol on the accessories after disinfection, for good maintenance can
extend the life of accessories.
Single-use accessories are not designed to be reused. Reuse may cause a risk of contamination and affect the
measurement accuracy.