Configuration Commands
On receipt of a configuration command, the anemometer will respond with a validation message if
the command was successfully completed, or an appropriate error message if it was not. If a
configuration command is entered with no data parameter, the current data value is returned.
Interrogation Commands
On receipt of an interrogation command, the anemometer responds with the requested information
if the command was successfully actioned, or an appropriate error message if it was not.
Invalid Combinations of Commands
It is possible through the setting of various parameters to request a configuration where the
required output bit rate exceeds the configured baud rate.
If an issued configuration command requests a parameter change that would cause the above
situation, an appropriate error message is generated.
13.2 Software Commands
|
Option, one option from a list separated by the vertical bar must be included in the
command.
()
Parenthesis, used to make the possible combinations unambiguous.
[]
Option, the symbols within the square braces are optional.
n{}m
Repeat, the symbols inside the braces can be repeated n to m times. If n is not
included then 0 is implied, if m is not present then infinity is implied.
<>
Non-literal symbol, this represents other non-literal and literal symbols. A
further definition is required to define this non-literal symbol.
Note:
Parameters in
bold
are the default settings.
?
Syntax:
? followed by sending CR/LF (pressing the Enter Key).
Description:
Request the current results message when in polled mode.
ABSTEMP
Syntax:
ABSTEMP OFF | (ON K | C)
Description:
Enables and disables absolute temperature reporting from the PRT sensor.
Parameters:
OFF
PRT temperature reporting turned off.
ON
PRT temperature reporting turned on.
K
PRT temperature reporting in K.
C
PRT temperature reporting in °C.
ALIGNUVW
Syntax:
ALIGNUVW AXIS | SPAR
Description:
Selects UVW axis alignment from two alternative definitions.
Parameters:
AXIS
Align the U axis to transducer axis 1 when viewed from above
.
SPAR Aligns the U axis with the North spar.
ANAIP
Syntax:
ANAIP <ipno>{,< ipno> }6 ( D | E )
<ipno> = 1 | 2 | 3 | 4 | 5 | 6
Description:
Turns on or off the specified analogue inputs.
Parameters:
ipno Analogue input channel number.
D
Disable specified channels.
E
Enable specified channels.
Example:
R3> ANAIP 1,3,5 E
R3> VANAIP
AOPFSD
Syntax:
AOPFSD 10 | 20 |
30
| 60
(Metres Per Second Only)
Description:
Specifies the analogue outputs full scale deflection. Used to configure the
analogue output box only.