HS-50 & HS-100
Doc. No 1199-PS-0035 Issue 3
June 2021
___________________________________________________________________________________________
Page 34
14 USING THE ANEMOMETER DIGITAL OUTPUTS
14.1 Digital Serial Output Format
Data is sent in either binary or ASCII format as follows:
14.1.1
Binary Result Message Format (post conversion)
01
10
176
00.1
00.0
350.4
+99.99
-0.0001
-0.0001
-0.0001
-0.0001
-0.0001
-0.0001
2B
StaA StaD Wc1 Wc2
Wc3
C
Abs T
An1
An2
An3
An4
An5
An6
cc
where:
0xBA 0xBA
-
Start of result message identifier. 2 bytes, hex value BA.
StaA
-
Status Address
StaD
-
Status Data
Wc 1
-
Wind component 1
Wc 2
-
Wind component 2
Wc 3
-
Wind component 3
C
-
Speed of Sound or Sonic Temperature
T
-
Absolute temperature
An
-
Analogue input 1
cc
-
checksum of all bytes. The checksum is an exclusive OR of all bytes
excluding the 2 start bytes.
These fields are defined as follows:
Wind Component Fields
UVW format
Wc1
U axis velocity in 0.01ms-1 units (Two's complement signed)
Wc2
V axis velocity in 0.01ms-1 units (Two's complement signed)
Wc3
W axis velocity in 0.01ms-1 units (Two's complement signed)
Polar format
Wc1
U V plane velocity direction in ° relative to N.
Wc2
U V plane velocity magnitude in 0.01ms-1 units.
W c3 W axis velocity in 0.01ms
-1 units (Two's complement signed)
Axis Speeds W c1 Ultrasonic axis 1 velocity in 0.01ms
-1 units (Two's complement
signed)
W c2 Ultrasonic axis 2 velocity in 0.01ms
-1 units (Two's complement
signed)
W c3 Ultrasonic axis 3 velocity in 0.01ms
-1 units (Two's complement
signed)
Speed of Sound Field
Speed Of Sound
Speed of sound in 0.01ms-1 units (Unsigned)
Sonic Temperature
Sonic temperature in 0.01K (Unsigned)
or 0.01°C (Two's complement signed)
Absolute Temperature Field
Absolute Temperature in 0.01K (Unsigned) or 0.01°C (Two's complement signed)
Analogue Input Fields
Analogue input value in twos complement:
0000
0.0000V
1FFF
+4.9994V
E000
-5.0000V
Checksum Field
Exclusive OR of all bytes excluding the two start bytes.