Weight calibration and linearization
With the instrument off, move jumper JP1 to the inside of the board (see installation manual)
Shutdown
instrument
CALibr PArAM
tESt
ConSt
SEr
bAtt
4 sec.
CApAC
Zero calibration
FS calibration
Set sample weight
During the calibration phase, the display shows the weight intermittently.
Calibration of
zero
Perform the operation with the scale unloaded but complete with the tare weight, with
the weight stabilized. The displayed weight should reset to zero. It is possible to repeat
this operation several times.
Calibration of
full scale
Before performing the operation, load the sample weight onto the balance and wait for
stabilization; the display will show the detected value to be calibrated.
If the set value is higher than the resolution offered by the instrument, it is not
is accepted and the display shows an error message for a few seconds.
It is always possible to repeat the calibration operations.
It is possible to cancel zero and full scale calibrations.
(long pressure):
cancels the zero calibration.
(long press):
Performs weight linearization (maximum 10 points)
Linearization procedure
CAL
Ln P1
Ln P2
Ln P10
4 sec.
Set sample weight 1st step
In programming the sample weight, values above the full scale, or below the
previous point, or at unstable weight.
If the value entered is accepted the next point is proposed otherwise still the same.
Linearization points are automatically reset to zero whenever the capacity is changed or if a full-scale
calibration is performed.
If you turn off the instrument without exiting the set-up menu, the
programming performed. When calibrations or settings are completed, return jumper JP1 to
the outside of the board.