28
Weighing parameter setting menu
With the instrument off, move jumper JP1 to the inside of the board (see installation manual) or hold down
the
key while pressing the
key.
Shutdown
instrument
PArAM SEr
ConSt CALibr
bAtt
tESt
4 sec.
Set value
Filt-
Set value
StAb-
Set value
Auto O
Set value
0 tr
Set value
G CAl
Only with jumper
Set value
G uSE
Only with jumper
Filt-
Digital weight filter factor
from 0 to 4 (default 2).
StAb-
Weight stability factor
0 to 4 (default 2).
Aut 0
Autozero threshold in weight value
(Default = 10% of range).
If autozero threshold = 0, autozero is not performed at power-up.
O tr
Automatic zero tracking factor
(0 to 4). (Default = 2).
The zero tracking factor determines the tracking speed. The maximum resettable value
corresponds to 2 % of the range.
G CAl
Gravity acceleration of the calibration site.
Gravity acceleration data are available only in the fiscal version with the calibration jumper
closed. Set only the decimal digits (e.g., G=9.79584, set 79584). Gravity acceleration
values are automatically reset, when the weight calibration is changed. Entering the value 0
cancels the correction on the original calibration
G uSE
Gravity acceleration of the place of use.
Gravity acceleration data are available only in the fiscal version with the calibration jumper
closed. Set only the decimal digits (e.g., G=9.79584, set 79584). Gravity acceleration
values are automatically reset, when the weight calibration is changed. Entering the value 0
cancels the correction on the original calibration