HD2010UC
- 64 -
V4.2
The STT:ACQUISITION commands are provided in the following table
Command
Format
Description
HOLD
D1
Interrupts display update
UPDATE
D1
Restarts display update
PAUSE
D1
Measurement in pause
RUN
D1
Starts measurements
STOP
D1
Ends measurements
CLEAR
D1
Clears measured levels
CONTINUE
D1
Restarts measuring
ERASE
D1
Erases the last x seconds of measure-
ments
RECORD
D1
Starts and records measurements
The STT:ACQUISITION:? command provides information on the acquisition status as shown in
the following example.
STT:ACQ:?
STT:ACQUISITION:STOP
BATTERY: 32%
MEMORY: 95.4%
DUMP TIME:00:00:01
LAST CALIBRATION: 2003/07/31 08:37
The STT:DISPLAY commands are listed in the table below.
Command
Format
Description
SLM
D1
Displays 5 selectable parameters in numeric format
PROFILE
D1
Displays the time profile of a selectable parameter
PROB_DISTR.
D1
Displays the level distribution of probabilities
CUMUL_DISTR.
D1
Displays the percentile levels graph
The STT:DISPLAY:? command provides information relevant to the sound level meter actual dis-
play as shown in this example.
STT:DIS:?
STT:DISPLAY:Mode:PROFILE
The following table lists the STT:MONITOR commands.
Command
Format
Description
ON
D1
Starts the Monitor function
OFF
D1
Ends the Monitor function
MEASUREMENT
D1
Monitor
SLM
D1
Monitor by 3 parameters
PROFILE
D1
Monitor by single parameter
REPORT
D1
Monitor of the reports
EVENT
D1
Monitor of the events