HD2010UC
- 14 -
V4.2
SLM
(
SOUND LEVEL METER
)
This is the display mode upon power on.
Three parameters (selectable among the following ones) can be displayed simultaneously:
•
Instantaneous
acoustic parameters such as L
p
, L
eq
(Short) and L
pk
. The instantaneous pressure
level is displayed as the maximum level reached every 0.5s.
•
Integrated
acoustic parameters with wideband frequency weightings, such as L
pmax
, L
eq,
L
Ieq
and L
pkmax
, updated every 0.5s.
•
Percentile levels selectable between L
1
ad L
99
.
•
Sound exposure level.
•
Daily personal exposure level.
•
Dose and daily Dose with programmable Exchange Rate, Criterion and Threshold Levels.
•
Overload Time (in %).
The display is updated every 0.5 seconds.
The data recording depends on the integration mode selected (single or multi) and on the activation
or not of the auto-store mode. (see the chapter “DESCRIPTION OF THE DIFFERENT
INTEGRATION MODES” on page 25).
Auto-Store: OFF
Auto-Store: ON
In
te
gr
at
ion
S
INGL
E
Recording twice a second of the 3 pa-
rameters and 8 times a second of LAF.
Automatic Stop at the end of the set inte-
gration interval.
Automatic recording of SLM page at the
end of the set integration interval.
M
U
LTI
PLE
Recording twice a second of the 3 pa-
rameters and 8 times a second of LAF.
Automatic reset of integrated levels at
every integration interval.
Automatic recording of the SLM screen at
intervals equals to the set integration time.
At the beginning of each period, integrated
levels and spectra are set to zero.
D
ISPLAY
D
ESCRIPTION
Left at the top of the display there are the recording status symbol and the overload or under-
range indicator (described at the beginning of this chapter). In the midst there is the integration in-
terval and on the right the acquisition time (hours:minutes:seconds). When the integration mode is
set on MULT (MENU>>General>>Integr.Mode:MULT), the “Tint” symbol flashes. The battery
symbol is in the right corner, indicating battery level.
56.9
60.8
79.3
30
110
01:08:25
Tint= s
10
LFp
Leq
Lpk
dBA
dBC
dBC
Integration interval
Acquisition time
Maximum level
Minimum level
Displayed parameters
Bar showing
istantaneous level