HD2010UC
- 24 -
V4.2
Per each event identified, 2010UC calculates the following:
•
5 programmable selectable parameters: maximum and minimum levels, peak level, equiva-
lent level and SEL.
•
Full statistical analysis
These parameters are not displayed but can be stored, completely or partially, at the end of each
level. The menu
Recording
>>
Event
allows selection of the 5 parameters and memorization.
The event parameters integration begins 2 seconds before triggering. This pre-trigger time cannot be
modified.
A special printing function, synchronized with the trigger, is available for communication of the
event itself via RS232 (Menu >> Trigger >> Print).
If the trigger is enabled, the sound level profile is displayed with black background when the
trigger is not active so as to highlight the event’s portion.
Fig. 9 - Description of the “Event trigger - display”