HD2010UC
- 91 -
V4.2
Press and hold down the ENTER key while reinserting the battery: the sound level meter will auto-
matically turn on. Confirm the load of factory calibration.
The sound level meter calibration parameters are restored to the last factory calibration; all menu
items are simultaneously brought back to the factory parameters (default).
M
ISCELLANEOUS
P
ROBLEMS
1.
The instrument does not turn on after battery replacement.
Remove one of the batteries and wait at least 5 minutes before reinserting it. The instru-
ment should turn on automatically when the battery is inserted.
2.
The detected sound levels seem incorrect.
Ensure that no condensation is present on the capsule or preamplifier. Avoid turning on
the instrument in conditions of possible condensation growth. Use the outdoor micro-
phone unit HD WME to perform measurements in conditions of high humidity or when it
is raining.
Check that the warm-up time, indicated by a blinking letter “W” over the upper left status
indicator, has elapsed.
Check measurement accuracy using the acoustic calibrator.
Load factory calibration.
Check that the microphone protection grid is screwed down on the capsule securely.
3.
Upon turning the sound level meter on, it turns off automatically right after the introduction
screen.
The batteries are flat.
4.
The sound level meter does not communicate with the PC.
Check that the sound level meter and PC communication speed is the same (MENU >>
Instrument >> Input/Output >> Baud Rate).
Check that the connection cable is plugged correctly in the sound level meter and is con-
nected to a PC serial port or USB with the item MENU >> Instrument >> Input/Output
>> Serial device set up respectively on RS232 or USB respectively.
If you use a USB interface, check that the driver is correctly installed.
If you use a Noise Studio program, disable the AutoDetect function (Menu Option >>
Port Settings) and set the connection directly to the COM port to which the instrument is
connected with the same baud rate of the sound level meter (Menu >> Instrument >> In-
put/Output >> Baud Rate).
5. It is not possible to activate continuous recording. By pressing REC and RUN, the instrument
starts measurements without recording.
The instrument does not have available memory for additional data. Download the data
and/or erase the memory.
The “Advanced Data Logger” option is not installed.