
HD2010UC
- 77 -
V4.2
Input Gain [dB]
Parameter
Lower Limit
[dB]
Upper Limit
[dB]
30
LpA
33
111
LpC
37
LpZ
40
LpkC
50
114
LpkZ
52
40
LpA
31
101
LpC
36
LpZ
40
LpkC
50
104
LpkZ
52
The starting level for the linearity range detection matches the reference level (94 dB) at 1
kHz. With different frequencies, the starting level takes into account the attenuation of the frequen-
cy weighting being measured. In the secondary fields the starting level is subject to the same incre-
ment of the input gain.
Integration Time
It can be set from a minimum of 1s to a maximum of 99 hours.
Measurement Dynamics with Electromagnetic Fields
Minimum measurable level equal to 60dBA with carrier from 26 MHz to 1 GHz and amplitude
equal to 10V/m modulated 80% at 1 kHz.
Reference Conditions
The measuring range is that with input gain equal to 10 dB.
The level corresponds to 94 dB.
The acoustic calibration can be performed at a sound level included in the range 94dB ÷
124dB.
The reference direction of the acoustic signal is that of the preamplifier longitudinal axis.
The reference acoustic field is the free field
Operating conditions
Storage temperature: -25 ÷ 70°C.
Operating temperature: -10 ÷ 50°C.
Working relative humidity: 25 ÷ 90%RH, not condensing.
Static pressure: 65 ÷ 108kPa.
Protection degree: IP64.
In case of condensation growth, it must evaporate completely before using the sound level
meter.
Drifts
Temperature: ± 0.3 dB over the range -10 ÷ 50°C (with correction for microphone drift acti-
vated).
Relative humidity: ± 0.3 dB over the range 25 ÷ 90%RH, not condensing.
Static pressure: ± 0.3 dB over the range 65 ÷ 108kPa.
Содержание DeltaOHM HD2010UC
Страница 2: ...HD2010UC 2 V4 2 ...
Страница 98: ...HD2010UC 98 V4 2 RIGHT Keypad The RIGHT key moves rightwards the cursor or the two active cursors flashing ...
Страница 129: ...HD2010UC 129 V4 2 NOTES ...
Страница 130: ...HD2010UC 130 V4 2 NOTES ...
Страница 131: ......