HD50
- 37 -
V1.5
Address
Type
Holding Register description
1044 +
200 x (
N
-1)
SW
Channel
N
lower alarm threshold in
mV
. Only if channel
N
is configured
as
0…10 V
input.
1045 +
200 x (
N
-1)
SW
Channel
N
higher alarm threshold in mV. Only if channel
N
is configured
as 0…10 V input.
1046 +
200 x (
N
-1)
SW
Lower alarm threshold expressed as value of the quantity associated to
channel
N
when the channel is configured as mapped 0…10 V input.
1047 +
200 x (
N
-1)
SW
Higher alarm threshold expressed as value of the quantity associated to
channel
N
when the channel is configured as mapped 0…10 V input.
1050 +
200 x (
N
-1)
SW
Channel
N
lower alarm threshold in
mV
(x100). Only if channel
N
is con-
figured as
-50…50 mV
input.
1051 +
200 x (
N
-1)
SW
Channel
N
higher alarm threshold in mV (x100). Only if channel
N
is
configured as
-50…50 mV input.
1052 +
200 x (
N
-1)
SW
Lower alarm threshold expressed as value of the quantity associated to
channel
N
when the channel is configured as mapped -50…50 mV.
1053 +
200 x (
N
-1)
SW
Higher alarm threshold expressed as value of the quantity associated to
channel
N
when the channel is configured as mapped -50…50 mV.
General information
da 10000
a 10019
B
User code with ASCII codification.
Acceptable values are in the set {32,…,126}.
10020
W
Current year
10021
W
Current month
10022
W
Current day
10023
W
Current hour
10024
W
Current minute
10025
W
Current second
10026
W
Measurement interval: 0=1s, 1=2s, 2=5s, 3=10s, 4=15s, 5=30s, 6=1min,
7=2min, 8=5min, 9=10min, 10=15min, 11=30min, 12=1h
10027
W
Logging interval: 0=1s, 1=2s, 2=5s, 3=10s, 4=15s, 5=30s, 6=1min, 7=2min,
8=5min, 9=10min, 10=15min, 11=30min, 12=1h
10036
W
Password to be supplied to enable configuration change commands. The
reading provides the fixed value 32768.
da 10037
a 10046
B
Device group with ASCII codification.
Acceptable values are in the set {32,…,126}.
10052
W
Setting of the quantities to be displayed in the automatic viewing cycle.
Set the i-th bit (starting from LSB) to 1 if you wish to include the i-th
quantity in the viewing cycle.
Example
: if in the model measuring and calculating: 1=Temp., 2=RH,
3=Td, 4=PVP, 5=Mix.Ratio, 6=AH, 7=Tw, the register is set to 0000 0000
0010 0010, only the relative humidity (RH) and the absolute humidity
(AH) will be displayed alternatively.
10053
W
Adding of the Wi-Fi signal level in the automatic viewing cycle.
Set LSB to 1 if you wish to include the Wi-Fi signal level in the viewing
cycle.
da 20000
a 20011
B
User code with ASCII codification of measurement #1.
da 20012
a 20023
B
User code with ASCII codification of measurement #2.
da 20024
a 20035
B
User code with ASCII codification of measurement #3.
da 20036
a 20047
B
User code with ASCII codification of measurement #4.