INTRODUCTION 

The data loggers of the 

HD50

 series allow indoor monitoring of various physical quanti-

ties. The data loggers are available for the monitoring of: 

 

Temperature 

 

Humidity 

 

Atmospheric pressure 

 

Differential pressure 

 

Carbon dioxide (CO

2

 

Illuminance 

 

Particulate Matter (PM1.0, PM2.5, PM4.0, PM10) 

The  models  that  measure  relative  humidity  and  temperature  calculate  absolute  hu-
midity, Dew Point, wet bulb temperature, mixing ratio and partial vapour pressure. 

Models  with  4  input  channels,  with  terminal  header  connections,  are  available  for  the 
connection of standard analog sensors: 

 

Transmitters  with  0÷20 or  4÷20 mA  current  output  and  -50÷50 mV,  0÷50 mV, 
0÷1 V or 0÷10 V voltage output 

 

Pt100 / Pt1000 temperature sensors and K, J, T, N, E type thermocouples 

 

Sensors with potentiometric output 

This allows extending the monitoring capability to countless other quantities, in addi-
tion to those listed above. 

The  versions  with  LCD  can  be  equipped  with  a 

custom

 

LCD 

(option 

L

)  or  with  a 

graphic

 

LCD 

(option 

G

). 

Three  LEDs  on  the  front  panel  indicate  the  status  of  power  supply,  LAN/WLAN  local 
network connection and alarm. 

The data loggers can be connected to a local network via the 

Wi-Fi

 or 

Ethernet

 inter-

face. The data logger allows the simultaneous operation of two communication proto-
cols: proprietary and 

Modbus TCP/IP

. The data logger manages up to 10 “TCP/IP Cli-

ent”  simultaneously.  If  the  local  network  is  connected  to  Internet,  the  data  can  be 
regularly sent to an 

FTP

 address, to the 

Cloud

 and via 

e-mail

For each detected quantity, two alarm thresholds can be set by the user. Exceeding a 
threshold is signaled acoustically, by means of the internal buzzer, visually, by lighting 
the alarm LED on the front panel, and remotely, by sending alarm 

e-mails

. An alarm 

hysteresis and a delay in the generation of the alarm can be configured for each detect-
ed quantity. 

Thanks to the integrated web server, you can configure the data logger and view the 
real time measurements from any PC, tablet or smartphone connected to the same lo-
cal network of the data logger by simply using a web browser. 

The 

HD35AP-S

 and 

HDServer1 

PC software allow configuring the data logger, down-

loading the data into a database and viewing the measurements. The HD35AP-S soft-
ware  allows  connecting  one  data  logger  at  a  time,  while  the  HDServer1  software  al-
lows  detecting  automatically  all  the  data  loggers  connected  to  the  network  and  the 
simultaneous connection to all the data loggers. The 

HD35AP-CFR21

 software option 

(available with both HD35AP-S and HDServer1) allows the protection of recorded data 
and configuration in response to 

FDA 21 CFR part 11

 recommendations. 

The internal clock can be regularly synchronized with a NIST reference server. 

Содержание Delta OHM HD50 Series

Страница 1: ...Operating manual Web Data Logger HD50 series Keep for future reference www deltaohm com English Companies Brands of GHM...

Страница 2: ...RATION 7 3 3 HD50GH CONNECTIONS 8 4 CONNECTION TO THE NETWORK 9 4 1 HD35AP CFR21 SOFTWARE OPTION 10 5 WEB SERVER 11 5 1 SETTINGS MENU 12 5 2 MONITOR MENU 18 5 3 CONNECTIVITY MENU 20 5 4 FILES MENU 27...

Страница 3: ...two communication proto cols proprietary and Modbus TCP IP The data logger manages up to 10 TCP IP Cli ent simultaneously If the local network is connected to Internet the data can be regularly sent...

Страница 4: ...or for illuminance probe only HD50 I models 6 CO2 fixed probe only HD50 B models 7 Differential pressure inputs only HD50 4r models The polarity is indicated next to the inputs 8 RJ45 connector for Et...

Страница 5: ...ing is active CLOUD symbol indicates that the data logger is configured to send the data to the Cloud It blinks until the connection with the server is established HDServer1 symbol indicates that the...

Страница 6: ...the fixed probe or the probe with cable the following indications are used TC Probe with cable M12 connector TV Fixed vertical probe without cable TCV Fixed sensors photometric probe with cable Availa...

Страница 7: ...ument logging parameters alarm thresholds quantities to be acquired network settings etc can be configured by connecting the instru ment to the PC via Ethernet or Wi Fi local network see chapter 4 and...

Страница 8: ...2 wires Pt100 Pt1000 3 wires Pt100 Pt1000 4 wires 0 50 mV 50 50 mV 0 1 V 0 10 V 0 20 mA 4 20 mA Thermocouple K J T N E Potentiometer The switches A and B are set to the right only in the configuration...

Страница 9: ...llows connecting one data logger at a time or HDServer1 it allows connecting several data loggers simultaneously application soft ware and the MySQL database management system included in the HD35AP S...

Страница 10: ...the instrument configuration and viewing of data in the database Each user can be assigned a different password for using the software There are also three levels of access Administrator Super user an...

Страница 11: ...S INFO Data logger general information CONFIGURATION Modbus address user code and belonging group MEASURES Selection of the quantities to be displayed LOGGING Logging settings ALARMS Alarm thresholds...

Страница 12: ...instrument o model serial number o user code o belonging group o Wi Fi signal level only if the Wi Fi interface is active o behaviour when the internal memory is full cyclic overwriting or not o logg...

Страница 13: ...ing interval is greater than the measuring interval the average of the measurements acquired during the logging interval will be stored The Erase Memory button allows erasing the data in the data logg...

Страница 14: ...hreshold is 10 C and the upper threshold is 60 C the hysteresis in C is equal to 60 10 x2 100 1 C The alarm is generated after the set delay time or immediately if the no delay op tion is selected If...

Страница 15: ...ts the quantities available to the left and the quantities to be displayed to the right The quantities are displayed in the order in which they appear in the list on the right Temperature unit of meas...

Страница 16: ...ltage input with associated physical quantity Mapped Voltage 0 50mV 0 50 mV voltage input with associated physical quantity Mapped Current 4 20mA 4 20 mA current input with associated physical quantit...

Страница 17: ...ity corresponding to the input value X1 e g 0 0 C X2 value of input signal in the second point e g 20 00 mA Y2 value of the physical quantity corresponding to the input value X2 e g 70 0 C x x 4 00mA...

Страница 18: ...format MEASURES panel Displays the current value of the quantities only those selected for logging The measurements in alarm are highlighted with a red background The L and H sym bols indicate whethe...

Страница 19: ...ual scroll interval X axis manual scroll buttons Pause Start automatic scroll Displays the data in the From To interval Single all quantities on the same graph Multiple a graph for each quantity Y axi...

Страница 20: ...of a data logger with IP address 192 168 1 2 in the web browser you need to type 192 168 1 2 40 The WLAN LAN module has two TCP IP ports For each of the two ports set the fol lowing o the protocol se...

Страница 21: ...obtain the address automatically dynamic IP address via DHCP protocol Dynamic Host Configuration Protocol Note it is suggested to use a static IP for the communication with the HD35AP S soft ware wit...

Страница 22: ...er the user name of the account o Account Password enter the password of the account o SMTP Server Address enter the outgoing mail server name supplied by your e mail service provider o SMTP Server Po...

Страница 23: ...a determined interval of time indicate the interval starting instant in the Start Date Time field and the interval ending instant in the Stop Date Time field then press Execute Note sending e mail an...

Страница 24: ...2 hours 4 hours 8 hours 12 hours 1 day 2 days 4 days 1 week o In the Data Mode field select the format of the data sent via FTP LOG format for the database and displaying with the Monitor function of...

Страница 25: ...od field The available options are Real Time 15 min 30 min 1 hour 2 hours 4 hours 8 hours 12 hours 1 day 2 days 4 days 1 week o To require the sending to the Cloud of all the data memorized subsequent...

Страница 26: ...ta memorized sub sequently to a given instant indicate the instant in the Date Time Upload field and press Execute o To require the sending to the HDServer1 software of all data memorized in a de term...

Страница 27: ...he files to be imported multiple files can be se lected The data loggers corresponding to the imported files appear in the Device list select a data logger and press View data to display the data CHAR...

Страница 28: ...ontent of the register is 184 the value is to be intended as 18 4 o x100 centesimal value expressed as an integer e g if the content of the reg ister is 500 the value is to be intended as 5 00 The com...

Страница 29: ...ower threshold alarm 2 higher threshold alarm 4 SW DEW POINT in the set measurement unit x10 5 B Dew Point alarm 0 OFF 1 lower threshold alarm 2 higher threshold alarm 6 SW PARTIAL VAPOR PRESSURE in h...

Страница 30: ...hreshold alarm 2 higher threshold alarm 184 SW TYPICAL PARTICLE SIZE in m x100 185 B Typical particle size 0 OFF 1 lower threshold alarm 2 higher threshold alarm Measured values and status of measurem...

Страница 31: ...annel N x10 Only if channel N is configured as 0 1 V input 1029 200 x N 1 B Alarm for channel N if the channel is configured as 0 1 V input 0 OFF 1 lower threshold alarm 2 higher threshold alarm 1030...

Страница 32: ...OFF 1 lower threshold alarm 2 higher threshold alarm 1052 200 x N 1 SW Value of quantity associated to channel N if the channel is configured as mapped 50 50 mV input 1053 200 x N 1 B Alarm for quanti...

Страница 33: ...red as mapped 0 50 mV see the table of indexes 6039 200 x N 1 SW Resolution of the quantity associated to channel N if the channel is config ured as mapped 0 50 mV 2 100 1 10 0 1 1 0 1 2 0 01 6040 200...

Страница 34: ...reshold in hPa x100 7 SW Partial vapor pressure higher alarm threshold in hPa x100 8 SW MIXING RATIO lower alarm threshold in g Kg x10 9 SW Mixing ratio higher alarm threshold in g Kg x10 10 SW ABSOLU...

Страница 35: ...1000 200 x N 1 SW Lower alarm threshold for TEMPERATURE with 2 wire Pt100 sensor of channel N in the set measurement unit x10 1001 200 x N 1 SW Higher alarm threshold for temperature with 2 wire Pt100...

Страница 36: ...1 200 x N 1 SW Channel N higher alarm threshold in mV x100 Only if channel N is configured as 0 50 mV input 1032 200 x N 1 SW Channel N lower alarm threshold in mA x100 Only if channel N is con figure...

Страница 37: ...020 W Current year 10021 W Current month 10022 W Current day 10023 W Current hour 10024 W Current minute 10025 W Current second 10026 W Measurement interval 0 1s 1 2s 2 5s 3 10s 4 15s 5 30s 6 1min 7 2...

Страница 38: ...83 B User code with ASCII codification of measurement 7 da 20084 a 20095 B User code with ASCII codification of measurement 8 da 20096 a 20107 B User code with ASCII codification of measurement 9 da 2...

Страница 39: ...Alarm Acoustic by means of the internal buzzer LED on the front pan el sending of e mails Power supply External 7 30 Vdc no internal battery PoE Power over Ethernet power supply via optional POE SPLT1...

Страница 40: ...TV Min 356 180 Max 906 640 T RH TD TW AH MR PVP P HD50PM Min 469 510 Max 906 640 PM1 0 PM2 5 PM4 0 PM10 TPS HD50 H Min 615 220 Max 1 165 680 depends on the connected sensors 1 Quantities T temperature...

Страница 41: ...ble units of measurement hPa mbar mmHg inchHg mmH2O inchH2O atm Differential pressure Sensor Piezoresistive Measuring range 125 Pa Resolution of the instrument 0 01 Pa Accuracy 0 35 typ of measuring s...

Страница 42: ...cordance with standard photopic curve V temperature coefficient f6 T 0 05 K Calibration uncertainty 4 f 1 accordance with photopic response V 6 f2 response as cosine law 3 f3 linearity 1 f4 instrument...

Страница 43: ...or coefficient 0 00385 C 1 Connection 2 3 or 4 wires Thermocouple Thermocouple type K J T N E The inputs are not isolated use thermocouples with isolated hot junction Measuring range type K 200 1370 C...

Страница 44: ...unscrew the probe filter screw the HD31 B3A adapter and connect the cylinder adjust the cylinder flow meter to get a constant flow between 0 3 and 0 5 l min CO2 calibration with a cylinder The calibr...

Страница 45: ...the specific measures described in this operating manual The proper operation and the operational safety of the instrument can be ensured on ly under the climatic conditions specified in this manual...

Страница 46: ...necting the power supply instruction manual The external probes and the SWD10M8 power supply or the POE SPLT12M8 PoE splitter or the CPM8 power supply cable have to be ordered separately The Ethernet...

Страница 47: ...uracy 0 3 C in the range 0 70 C 0 4 C outside Diameter 14 mm 4 pole M12 female connector TP35 1 Temperature probe with Pt1000 1 3 DIN 4 wire sensor Operating tem perature 50 105 C Dimensions 5 x 40 mm...

Страница 48: ...the HP3517 probe requires recalibration of the instrument in line with the new probe HD9007A 1 12 ring protection against solar radiations Includes support bracket HD9007A 2 16 ring protection agains...

Страница 49: ...r HD75 Saturated solution to check Relative Humidity probes at 75 RH includes ring adapter for 14 mm diameter probes thread M12 1 HD33 Saturated solution to check Relative Humidity probes at 33 RH inc...

Страница 50: ...ent from that to which the receiver is connected Consult the dealer or an experienced radio TV technician for help Notice To satisfy FCC IC RF exposure requirements for mobile and base station transmi...

Страница 51: ......

Страница 52: ...ized third parties is excluded from the guarantee The instrument shall be returned FREE OF SHIPMENT CHARGES to your dealer The jurisdiction of Padua applies in any dispute The electrical and electroni...

Страница 53: ...e result of the constant development of the product This may produce some differences between the information written in this manual and the instrument you have purchased We cannot completely exclude...

Страница 54: ...GHM GROUP Delta OHM Delta Ohm S r l a socio unico Via Marconi 5 35030 Caselle di Selvazzano Padova ITALY Phone 39 049 8977150 Fax 39 049 635596 www deltaohm com info deltaohm com V1 5 08 04 2020...

Отзывы: