HD50
- 29 -
V1.5
Coils
- Read/Write parameters
Address
Type
Coil description
1
b
Logging status: 0=active, 1=inactive
2
b
Logging mode: 0=non cyclic, 1=cyclic
3
b
Set 1 to delete the device logging memory. Bit zeroing is automatic.
4
b
Buzzer activation in case of measurement alarm: 0=no, 1=yes
9
b
Protection of configuration with password: 0=no, 1=yes
Changing the parameter requires the Administrator password (see Holding
Register 10036).
71
b
Type of calibration in use: 0=factory, 1=user
72
b
Set 1 to perform the differential pressure calibration at zero. Bit zeroing
is automatic.
73
b
Differential pressure self-calibration status: 0=off, 1=on
Input Registers
- Read-only parameters
Address
Type
Input Register description
Measured values and status of measurement alarms
0
SW
T
EMPERATURE
with
NTC10K
sensor of channel
1
in the set measurement
unit (x10).
1
B
Alarm for temperature with NTC10K sensor of channel 1:
0=OFF, 1= lower threshold alarm, 2= higher threshold alarm
2
SW
R
ELATIVE
H
UMIDITY
in % (x10).
3
B
Relative humidity alarm:
0=OFF, 1=lower threshold alarm, 2=higher threshold alarm.
4
SW
D
EW
P
OINT
in the set measurement unit (x10).
5
B
Dew Point alarm:
0=OFF, 1=lower threshold alarm, 2=higher threshold alarm.
6
SW
P
ARTIAL VAPOR PRESSURE
in hPa (x100).
7
B
Partial vapor pressure alarm:
0=OFF, 1=lower threshold alarm, 2=higher threshold alarm.
8
SW
M
IXING
R
ATIO
in g/Kg (x10).
9
B
Mixing ratio alarm:
0=OFF, 1=lower threshold alarm, 2=higher threshold alarm.
10
SW
A
BSOLUTE
H
UMIDITY
in g/m
3
(x10).
11
B
Absolute humidity alarm:
0=OFF, 1=lower threshold alarm, 2=higher threshold alarm.
12
SW
W
ET
B
ULB TEMPERATURE
in the set measurement unit (x10).
13
B
Wet bulb temperature alarm:
0=OFF, 1=lower threshold alarm, 2=higher threshold alarm.
16
SW
T
EMPERATURE
with
NTC10K
sensor of channel
2
in the set measurement
unit (x10).
17
B
Alarm for temperature with NTC10K sensor of channel 2:
0=OFF, 1=lower threshold alarm, 2=higher threshold alarm.
20
SW
I
LLUMINANCE
in lux (low range, models HD50...I...).
21
B
Illuminance (low range, models HD50...I...) alarm:
0=OFF, 1=lower threshold alarm, 2=higher threshold alarm.
24
SW
A
TMOSPHERIC
P
RESSURE
in the set measurement unit (the multiplier de-
pends on the set unit).
25
B
Atmospheric pressure alarm:
0=OFF, 1=lower threshold alarm, 2=higher threshold alarm.
32
SW
CO
2
in ppm.