Содержание GL350

Страница 1: ...midi LOGGER SERVICE MANUAL 450 400 350 GL400 UM 251 GL400 UM 251 08 9370...

Страница 2: ......

Страница 3: ...ber for the input terminal units added 6 1 03 4 04 10 15 50 Ch terminal unit information added 4 5 4 7 5 2 5 7 5 8 03 5 04 10 22 Some page couldn t open It corrected 3 1 04 6 04 10 28 Some problem in...

Страница 4: ...vice personnel and other persons who have undergone equivalent technical training Maintenance or repair by unauthorized service personnel is to be strictly avoided Because the GL450 400 350 contains n...

Страница 5: ...the System Setup Menu 4 1 4 2 Setting Up Setup1 Voltage Input 4 2 4 3 Setup2 Room Temperature Compensation 4 5 4 4 Setting Up Setup3 RTD Temp 4 8 5 INSPECTION AND CHECK PROCEDURES 5 1 5 1 Insulation...

Страница 6: ...GL400 UM 251 9370 iv...

Страница 7: ...channels terminal unit 50SU 50 channels unit 2 units mountable for a maximum of 100 channels Battery Pack B 511 Battery Charger B 512 Logic Alarm Cable B 513 DC Drive Cable B 514 Cable to connect main...

Страница 8: ...GL400 UM 251 9370 1 2 1 CONFIGURATION...

Страница 9: ...t the monitor contrast 3 Copy the two main program les to a PCMCIA card from the computer How to upgrade 1 Insert the PCMCIA card containing the main upgrade program into the GL450 400 350 s PCMCIA ca...

Страница 10: ...LAN Cable How to upgrade OS Windows 2000 XP Software GL40upd exe Upgrade software le xxx dll Upgrade software dll le Firmware Mxxxa001 G40 Main Program le for GL450 400 350 Mxxxa002 G40 Main Program...

Страница 11: ...GL400 UM 251 9370 2 3 2 UPGRADING THE FIRMWARE 5 Click the Open File icon to show the menu as below Open File Icon 6 Select the main update le as Mxxxa001 G40 7 Click the Open Button...

Страница 12: ...lso your computer s TCP IP port setting must be set to the same segment as 192 168 0 X 9 Click the Start button to start upgrading when you are ready to upgrade GL450 400 The following message will ap...

Страница 13: ...GL400 UM 251 9370 2 5 2 UPGRADING THE FIRMWARE...

Страница 14: ...tion to avoid an electrical shock When handling the LCD shield and LCD protective plate be careful not to leave ngerprints or other marks on them 3 2 Mounting and Removing the Input Terminal Unit This...

Страница 15: ...ounting the Input Terminal Unit 1 Remove the cover from the input terminal mounting area 2 As shown in the gure below insert the input terminal unit in the GL450 400 350 At this time be sure to check...

Страница 16: ...er 3 Release the top cover hook from the bottom case and then remove the top cover assembly GL450 400 or the top cover GL350 NOTE Take care with the exible cable connecting the top cover assembly and...

Страница 17: ...GL400 UM 251 9370 3 4 3 DISASSEMBLY AND REASSEMBLY GL350 Hook Top cover Bottom case assembly M3L30 binding head screw M3L10 self tapping screw...

Страница 18: ...the main board from the bottom case assembly Insert this side first when installing the main board to the bottom case assembly Spacer for GL350 M2 6L6 binding head screw Main board Re assembly 1 Re as...

Страница 19: ...Section 3 2 2 Remove the main board from the bottom case See Section 3 3 3 Remove the four M2 6L6 binding head screws holding the AMP shield plate and the AMP board and then detach the AMP board from...

Страница 20: ...ottom case assembly AMP board M2 6L6 bindimg head screw AMP shield plate AMP sheet Fit the AMP sheet edge to the AMP board edge at these locations 2 Tighten the top two M2 6 binding head screws as sho...

Страница 21: ...ld plate and then tighten the bottom two M2 6 binding head screws as shown below Bottom M2 6 binding head screw AMP shield plate 4 Re assemble the other parts in the reverse order in which they were d...

Страница 22: ...the input terminal unit from the GL450 400 See Section 3 2 2 Disconnect the LCD exible cable from the control panel board Top cover assembly Control panel board LCD LCD flexible cable 3 Remove the on...

Страница 23: ...GL400 UM 251 9370 3 10 3 DISASSEMBLY AND REASSEMBLY Re assembly 1 Re assemble the LCD unit in the reverse order in which it was disassembled...

Страница 24: ...he UP Direction key 10 times in order to display the SET UP menu shown below 3 Move the cursor to the Mode position using the Direction keys 4 Press the ENTER key to display the menu below from which...

Страница 25: ...up to 50 VDC four digit output Mutually connect the ground terminals of the voltage generator and the GL450 400 350 Wiring connection 0 00000 V Voltage generator FG GND AC adapter Connect to CH1 Inpu...

Страница 26: ...the AMP Setup conditions as shown below Mode Setup1 Basic Slot Slot1 CH CH 1 4 Move the cursor to the Execute Setup position as shown above 5 Press the ENTER key to start the Setup1 Basic operation Me...

Страница 27: ...V 500 mV 500 000 mV 1 V 1 000 V 2 V 2 000 V 5 V 5 000 V 10 V 10 000 V 20 V 20 000 V 50 V 50 000 V 9 Repeat step 8 by sequentially setting the Measure parameter to 50 mV 100 mV 200 mV 500 mV 2 V 5 V 10...

Страница 28: ...ture device Zero Controller Connect CH3 and CH8 for the 10 channel input terminal unit Connect CH3 CH8 CH13 and CH18 for the 20 channel input terminal unit Connect CH3 CH8 CH13 CH18 CH23 CH28 CH33 CH3...

Страница 29: ...the cursor to the Execute Setup position as shown above 5 Press the ENTER key to start the Setup2 Room temp operation Message 6 The rTemp START message is displayed at the location shown above 7 Press...

Страница 30: ...l unit repeat steps 4 to 8 until Area 5 of Sensor2 14 Turn off the GL450 400 350 This completes the Room temp setup Setup ow for the Room temp Procedure 10 channel input terminal 20 channel input term...

Страница 31: ...on As shown below use a resistor box Wiring connection AC adapter Resistor Box Connect to and b terminals of CH2 Connect one of the resistor terminals and GND for the 0 ohm setting Connect both of the...

Страница 32: ...fy the AMP Setup conditions as shown below Mode Setup3 RTD temp Slot Slot1 CH CH 2 4 Move the cursor to the Execute Setup position as shown above 5 Press the ENTER key to start the Setup3 RTD temp ope...

Страница 33: ...you have nished the 350 ohm Setup operation the EEPROM ENTER message appears Press the ENTER key to begin writing the new Setup data to the EEPROM 13 After the setup parameters have been registered to...

Страница 34: ...sulation resistance gauge to check that the insulation resistance between the following terminals is equal to or higher than the rated value Measuring Terminals Rated value Remark Ch1 b to GND FG Mego...

Страница 35: ...Condition settings for the K range accuracy for the 10 Ch and 20 Ch terminal unit CondK50 1 CND Condition settings for the K range accuracy for ch3 8 13 18 23 of the 50 Ch terminal unit CondK50 2 CND...

Страница 36: ...S 3 Press the ENTER key to display the menu below 4 Specify the condition settings le to execute the various inspection operations 5 Move the cursor to the Execute position using the Direction keys 6...

Страница 37: ...nect the ground terminals of the voltage generator and the GL450 400 350 prior to initiating measurement Wiring connection Same wiring connection as Setup1 See Setup1 procedure Setting procedure 1 Loa...

Страница 38: ...tings Range Input voltage Rating 20 mV 20 000 mV 19 984 mV to 20 016 mV 50 mV 50 000 mV 49 96 mV to 50 04 mV 100 mV 100 00 mV 99 92 mV to 100 08 mV 200 mV 200 00 mV 199 84 mV to 200 16 mV 500 mV 500 0...

Страница 39: ...ayed and then display the values as shown below Message Con rm that this value is within the above ratings after data has been captured The data with an automatically generated le name is captured to...

Страница 40: ...specify output up to 1 V Wiring connection AC adapter T compensation lead T input side T output side 0 C reference temperature device Zero Controller Connect CH3 and CH8 for the 10 channel input term...

Страница 41: ...ngs le for type T as described above 2 Connect to the zero controller with the T compensation leads 3 Use the voltage generator to successively input the reference voltage signal for each range to the...

Страница 42: ...ettings le CondRTD CND as described above 2 As shown in the gure above connect the output of the resistor box to the GL450 400 350 s input terminal 3 Use the resistor box to successively set the refer...

Страница 43: ...bove 2 Press the DISPLAY key until the menu shown below appears 3 Press the START key to start data capture 4 Wait until the Finished message is displayed and then display the values as shown below Me...

Страница 44: ...320 20 ch Input Terminal Unit 20SU 1 or 2 GL400 350 795350370 50 ch Input Terminal Unit 50SU 1 or 2 GL400 350 795401110 10 ch Input Terminal Unit 10TU 1 or 2 GL450 795401120 20 ch Input Terminal Unit...

Страница 45: ...me Description Q ty Remarks 1 604400011 Rubber Grip 1 2 604350950 Emblem Plate GRAPHTEC EFBL 40 1 GL400 604450040 Emblem Plate 450 GRAPHTEC EFBL 40 GL450 1 GL450 3 604400020 Top Cover for GL400 Case t...

Страница 46: ...on Q ty Remarks 1 604400020 Top Cover for GL400 Case top LCD 1 GL400 604450060 Top Cover for GL450 Case top 450 LCD 1 GL450 2 604400004 Key Switch 1 3 770435105 Control Panel Board 1 GL400 795401701 C...

Страница 47: ...Name Description Q ty Remarks 1 690440031 Cable GL400 03 1 2 690440040 Cable GL400 04 1 3 500052325 Inverter CXA K0505 VJL 1 4 690440020 Flexible Cable GL400 02 1 5 604400070 Inverter Insulating Shee...

Страница 48: ...251 9370 6 5 6 PARTS LISTS 6 2 4 Top Cover GL350 No Part No Part Name Description Q ty Remarks 1 604350950 Emblem Plate GRAPHTEC EFBL 40 1 2 604350040 Top Cover for GL350 Case top normal 1 Top Cover G...

Страница 49: ...river Spring 1 5 604350180 AMP Sheet 1 6 604400090 Rear Lift up Hook for GL400 1 GL450 400 604350280 Rear Lift up Hook for GL350 1 GL350 7 770435103 AMP Board 1 8 604350151 AMP Shield Sheet 1 9 604400...

Страница 50: ...l Japanese model only 1 Japanese only 4 054025740 Serial Number Label 1 5 604350060 Bottom Case 1 6 795350310 Input Terminal Unit 10SU 1 GL400 350 Input Terminal Unit 10TU 1 GL450 7 604350941 Model Na...

Страница 51: ...l Number Label 1 5 770435108 Input Terminal Board 1 6 770435109 Input Terminal Relay Board 1 7 604350010 Input Terminal Top Case 1 10SU 604450100 Input Terminal Top Case 450 1 10TU 8 604350030 Input T...

Страница 52: ...AT Label 1 5 604350240 20 CH Insulating Sheet 1 6 054025740 Serial Number Label 1 7 604350030 Input Terminal Bottom Cover 1 20SU 604450110 Input Terminal Bottom Cover 450 1 20TU 8 604350112 Input Term...

Страница 53: ...light GL450 400 350 Block Diagram Battery Pack Option Battery Relay Board Power Switch AC Adapter AMP Board LCD Main Control Board Input Terminal Board Relay Board PCMCIA Card USB Emulator Flash LAN 5...

Страница 54: ...P7 P11 P4 P4 To KEY Board P15 To Humidity Sensor GL450 only To AMP Board Main Control Board GL400 01 AC Adapter AC IN Primary circuit Primary circuit Signal input circuit ISOLATING Connector to Conne...

Страница 55: ...9 PCMCIA Slot P11 LCD P3 P2 P1 FC GL400 04 FC GL400 01 CA GL400 03 BACK LIGHT Main Control Board GL400 01 LOGIC BATTERY P6 P2 P12 P15 P10 P8 LAN USB 5V Output GL450 only REMOTE P1 Flash Write Board BA...

Страница 56: ...ut circuit Signal Input circuit Signal Input circuit Signal Input circuit Signal Input circuit Signal Input circuit Signal Input circuit Signal Input circuit Signal Input circuit Signal Input circuit...

Страница 57: ...t circuit Signal Input circuit Signal Input circuit Signal Input circuit Signal Input circuit Signal Input circuit Signal Input circuit Signal Input circuit Signal Input circuit Signal Input circuit S...

Страница 58: ...RELAY Constant current Attenuator Selector Selector A D AMP AMP AMP AMP DC 5V DC 5V DC 24V Isolating Isolating Isolating DC DC ICL7660 DC DC OEJ05SC0512 Secondary circuit Main Control Board Photocoupl...