HD2328
- 16 -
V1.7
INSTRUMENT
TECHNICAL
CHARACTERISTICS
Instrument
Dimensions (Length x Width x Height)
140 x 88 x 38 mm
Weight
160 g (complete with batteries)
Material
ABS
Display
2x4½ digits plus symbols
Visible area: 52 x 42 mm
Operating conditions
Operating temperature
-5…+50°C
Warehouse temperature
-25…+65°C
Working relative humidity
0…90% RH without condensation
Protection degree
IP67
Power
Batteries
3 1.5V type AA batteries
Autonomy
200 hours with 1800mAh alkaline
batteries
Power absorbed with instrument off
< 20
µ
A
Connections
Inputs for probes
2-pole female polarized standard
miniature connectors
Unit of Measurement
°C - °F;
Measurement of temperature by Instrument
Tc measurement range: K
-200 … +1370°C
Tc measurement range: J
-100 … +750°C
Tc measurement range: T
-200 … +400°C
Tc measurement range: E
-200 … +750°C
Resolution
0.1°C
Instrument accuracy
Thermocouple K
±0.1 °C up to 600 °C
±0.2 °C over 600 °C
Thermocouple J
±0.1 °C up to 400 °C
±0.2 °C over 400 °C
Thermocouple T
±0.1 °C
Thermocouple E
±0.1 °C up to 300 °C
±0.2 °C over 300 °C
The precision only refers to the instrument. Error due to the thermocouple or
to the cold junction reference sensor is not included.
Temperature drift @20°C
0.02%/°C
Drift after 1 year
0.1°C/year
Accuracy of the thermocouple probes:
Содержание Delta OHM HD2328.0
Страница 4: ...HD2328 4 V1 7 Thermocouple Thermometer HD2328 0 ...
Страница 19: ......