HD2328

 

-   11   -

 

V1.7

 

The calibration information is saved in the instrument's memory. The same correction 

is applied to any probe that is subsequently connected to the input: therefore,  the 

"USER calibration" should only be used with the probe used for calibration and not 

with other probes. 

 

To pass from the user to the factory calibration and back, proceed as follows: 

 

 

press simultaneously (A-B)/MENU and DATA/ENTER to open the menu: 

 

press ENTER until the menu item "CAL_MODE" is selected. 

 

use the 

 and 

 arrows (respectively located above the HOLD and REL keys) to 

select the type of calibration, FACT or USER; 

 

confirm by pressing DATA/ENTER

 

C

ALIBRATION SEQUENCE 

 

I

NSTRUMENT ON LINE WITH THE PROBE

(

S

): 

The calibration can be carried out on one or two points that should differ by at 

least 10°C and be included in the probe functioning range. 

 

1.

 

Insert the probe into a thermostatic bath, the temperature of which is precisely 

known from a reading taken on a sample reference thermometer. Wait for the 

measurement to stabilize. 

2.

 

Press simultaneously (A-B)/MENU and DATA/ENTER to open the menu. 

3.

 

Press DATA/ENTER until the menu item "CAL_MODE" is selected. 

4.

 

Use the 

 and 

 arrows (respectively located above the HOLD and REL keys) to 

select the USER calibration. 

5.

 

Confirm with DATA/ENTER.  

6.

 

Use the 

 and 

 arrows (respectively located above the HOLD and REL keys) to 

select the input to which the probe being calibrated is connected, choosing the 

upper connector A or the lower one B. 

7.

 

Confirm with DATA/ENTER

8.

 

The "SEL_MEAS_1/2" message is displayed in the comment line. 

9.

 

Use the 

 and 

 arrows to select "1" (first calibration point). 

10.Confirm with DATA/ENTER

11.The  "UP DOWN 1st MEAS" message is displayed in the comment line: the 

instrument display shows the measured temperature. 

12.Use the 

 and 

 arrows to correct the indicated value until it coincides with the 

value measured by the sample reference thermometer. 

13.Confirm with DATA/ENTER

14.To exit the procedure without performing the second point, select "0" and press 

ENTER

15.To calibrate the second point, select the point "2" using the 

 and 

 arrows. 

16.Press DATA/ENTER

17.The "UP DOWN 2nd MEAS" message is displayed in the comment line. 

18.Move the probe to the second thermostatic bath and wait for the measurement to 

stabilize. 

19.The instrument display shows the measured temperature. 

20.Use the 

 and 

 arrows to correct the indicated value until it coincides with the 

value measured by the sample reference thermometer. 

21.Confirm with DATA/ENTER

  

The procedure is now complete. 

Содержание Delta OHM HD2328.0

Страница 1: ...Operating manual Thermocouple Thermometer HD2328 0 www deltaohm com Companies Brands of GHM English Keep for future reference ...

Страница 2: ...LIBRATION OF THE INSTRUMENT ON LINE WITH THE PROBE S 10 CALIBRATION SEQUENCE INSTRUMENT ON LINE WITH THE PROBE S 11 WARNINGS 12 INSTRUMENT SIGNALS AND FAULTS 13 INSTRUMENT STORAGE 13 LOW BATTERY WARNING AND BATTERY REPLACEMENT 14 WARNING ABOUT BATTERY USE 14 NOTES ABOUT WORKING AND OPERATIVE SAFETY 15 INSTRUMENT TECHNICAL CHARACTERISTICS 16 ORDERING CODES 18 ...

Страница 3: ...e chosen among the following types Type K Type J Type T Type E The instrument accepts the input of one or two thermocouples of the same type K J T E even in a different form The units of measurement are the following 1 C Celsius degrees 2 F Fahrenheit degrees Using the Max Min and Avg function of this instrument respectively obtains also the maximum minimum or average values Other available functi...

Страница 4: ...HD2328 4 V1 7 Thermocouple Thermometer HD2328 0 ...

Страница 5: ...nt display during normal operation in the menu increases the current value 8 A B MENU key during normal operation it displays in the secondary line the difference of the temperatures measured by the probes connected to the inputs A and B when pressed together with the DATA key it allows to open the menu 9 REL key during normal operation enables the relative measurement displays the difference betw...

Страница 6: ...ctions ON OFF to turn the instrument on press ON to turn it off press OFF The turning on enables all display segments for a few seconds and then the type of calibration enabled CAL FACT factory calibration CAL USER user calibration AUTO OFF the AutoPowerOff function can be disabled by simultaneously pressing this key and the HOLD key when turning the instrument on If no probe is connected to one o...

Страница 7: ...until the FUNC_CLRD message appears ENTER once the MENU has been opened with the DATA UNIT keys the DATA key will perform the ENTER function and the MENU can be browsed and the displayed parameter confirmed The CLR key has two functions CLEAR CLR allows to reset the maximum MAX minimum MIN and average AVG value of the captured measurements ESC once the MENU has been opened with the DATA UNIT keys ...

Страница 8: ...multaneously DATA A B the first item of the instrument programming menu will appear use the and arrows respectively located above the HOLD and REL keys to modify the displayed value press DATA ENTER to confirm the modification and go onto the next item press CLR ESC to cancel the modification to exit the menu press the A B MENU key again The REL key is used for the following functions REL it displ...

Страница 9: ...t item press CLR ESC to cancel the modification to exit the menu press the A B MENU key again 3 Starting the User calibration procedure the _CAL_MODE message is displayed in the comment line and FACt is displayed in the main line The instrument is provided with the factory FACt calibration It is also possible to perform a USER calibration USEr of instrument probe The calibration information is sav...

Страница 10: ...re measured by immersion is carried out by inserting the probe in the liquid in the oven where you wish to perform the measurement the hot junction is housed in the end part of the probe In the temperature measured by penetration the probe tip must be inserted into the material the hot junction is housed in the end part of the probe NOTE when measuring the temperature on frozen blocks it is conven...

Страница 11: ...d arrows respectively located above the HOLD and REL keys to select the USER calibration 5 Confirm with DATA ENTER 6 Use the and arrows respectively located above the HOLD and REL keys to select the input to which the probe being calibrated is connected choosing the upper connector A or the lower one B 7 Confirm with DATA ENTER 8 The SEL_MEAS_1 2 message is displayed in the comment line 9 Use the ...

Страница 12: ...s for surface measurements contact probes must be held perpendicular against the surface Apply oil or heat conductive paste between the surface and the probe in order to improve contact and reduce reading time Whatever you do do not use water or solvent for this purpose A contact measurement is always very hard to perform It has high levels of uncertainty and depends on the ability of the operator...

Страница 13: ...is not connected FUNC CLRD Max min and average values clearing OVER o UNDR Measurement overflow indicates that the probe is measuring a value exceeding the measuring range PLS_EXIT FUNC RES_FOR_FACT ONLY Please exit using ESC function reserved to factory calibration SEL CHAN Input channel selection SEL MEAS 1 2 Select the first second calibration point SEAL MEAS UNIT Selection of the unit of measu...

Страница 14: ...n the instrument back on To replace the batteries proceed as follows 1 switch the instrument off 2 unscrew the battery cover counter clockwise 3 replace the batteries 3 1 5V alkaline batteries type AA 4 screw the cover on clockwise Malfunctioning upon turning on after battery replacement After replacing the batteries the instrument may not restart correctly in this case repeat the operation After ...

Страница 15: ...ional safety of the instrument can only be guaranteed under the environmental and electrical operating conditions that are in specified in chapter TECHNICAL CHARACTERISTICS Do not use or store the product in places such as listed below Rapid changes in ambient temperature which may cause condensation Corrosive or inflammable gases Direct vibration or shock to the instrument Excessive induction noi...

Страница 16: ...onnections Inputs for probes 2 pole female polarized standard miniature connectors Unit of Measurement C F Measurement of temperature by Instrument Tc measurement range K 200 1370 C Tc measurement range J 100 750 C Tc measurement range T 200 400 C Tc measurement range E 200 750 C Resolution 0 1 C Instrument accuracy Thermocouple K 0 1 C up to 600 C 0 2 C over 600 C Thermocouple J 0 1 C up to 400 C...

Страница 17: ...refer to the expected thermocouple operating temperature in agreement with the thermoelements diameter Tolerance of the thermocouples Type of themocouple Range C G I G II K 0 1370 C 1 1 C or 0 4 2 2 C or 0 75 J 0 750 C 1 1 C or 0 4 2 2 C or 0 75 T 0 400 C 0 5 C or 0 4 1 C or 0 75 E 0 750 C 1 C or 0 4 1 7 C or 0 5 K 200 0 C 2 2 C or 2 T 200 0 C 1 C or 1 5 E 200 0 C 1 7 C or 1 The higher of the two ...

Страница 18: ...UPLE PROBES The thermocouple probes can be connected to all instruments using the standard miniature connector which can be obtained from the price list DELTA OHM metrology laboratories LAT N 124 are ISO IEC 17025 accredited by ACCREDIA for Temperature Humidity Pressure Photometry Radiometry Acoustics and Air Velocity They can supply calibration certificates for the accredited quantities ...

Страница 19: ......

Страница 20: ...warranty included in the manual of the product For any dispute the competent court is the Court of Padua The Italian law and the Convention on Contracts for the International Sales of Goods apply TECHNICAL INFORMATION The quality level of our instruments is the result of the continuous product development This may lead to differences between the information reported in the manual and the instrumen...

Отзывы: