Instruction Manual
ION Director®
EN
2021-01-23
21
Assembly stacked and next to each other
Please note that depending on the chosen mounting method you need different tube lengths
between
IOND
and
Doser
. For orientation, you will find below a tabular overview of the
recommended tube lengths.
TIP
Make a note of the assembled tube length for taking the water sample
(here pump 1) and the tube length to the
Water Sample
input of the ION
Director® and add the two lengths. You will need this total tube length
later when determining the
Sample tube volume
.
In order to achieve maximum measuring accuracy, it is necessary to keep the tube to the
reference liquids as short as possible (in total max. 100 cm / 40
”
).
Please make sure that bubbles cannot collect anywhere in the tube, for example because the
tube sags.
Assembly
Water sample
Reference A
Reference B
stacked
16 cm (6.3“)
17 cm (6.7“)
18 cm (7.08
“)
side by side
26 cm (10.2“)
23,5 cm (9.3“)
21 cm (8.3“)