17
IIX – Periodic Preventive Maintenance
A-
Water checking list
Monthly check the filters cleanness.
Check the hot water: Between 50° & 65° C (Neutral Ph. 7°).
Check the Cold Water temp: Between 5° & 15°C (Neutral Ph. 7°).
Check the Service Pressure: Between Min 0.8 Bars Max 2.00Bars.
Check Hot and Cold Water hardness to be the nearest to neutral = Ph. 7° / + -2°.
Check any leaks on all the piping installation.
B-
Electricity checking list
Monthly verify electricity power supply voltage corrects range
single phase 120V/60Hz.
Check GFCI reaction.
Check the external equipotential electric main security short circuit breaker and check your equipotential insulation.
Check the electricity stability.
Check all line upstream GFCI Breakers
IX-
Environmental
conditions for storage and transport
Condition
Value
Temperature
Ambient temperature -10°C to 50°C / 14°F to 122°F
Air pressure
700 to 1060 hPa
Relative humidity
20% to 90% @ 30°C, non-condensing
Ambient conditions during use
Condition
Value
Temperature
Ambient tempe5°C to 40°C 14°F to 122°F
Air pressure
700 to 1060 hPa
Relative humidity
20% to 75% @ 30°C, non-condensing
Содержание Pedispa Compact
Страница 6: ...6 Unfolded position fig 2 Full Spread Position fig 3...
Страница 7: ...7 III Before Installation 1 Over All Dimension Over All Dimension front view fig 4 Metric Inch...
Страница 13: ...13 Perspective of the Base frame fixation and inlets implementation drawing fig 11...
Страница 19: ...19 Annex 1 Space requirements implementation example drawings Corner space Implementation example fig 16...
Страница 20: ...20 Room space dual Pedispa compact Implementation example fig 17...
Страница 21: ...21 Room space Quadruple Pedispa compact Implementation example fig 18...
Страница 22: ...22 Room space Pedispa compact Circular Implementation Centralized single backwash station example fig 19...