4
PeakStore5 Manual Version 3.0
1.
General
The data acquisition system PeakStore5 is used for data acquisition on mechanical drives for
vibration diagnostics purposes. The PeakStore5 is available in 3 different variants:
•
PeakStore504 with 4 input channels
•
PeakStore508 with 8 input channels
•
PeakStore512 with 12 input channels
It can be connected with accelerometers with constant current feed, inductive displacement
sensors or other isolated signals with ±10 V. The data is measured with up to 51.2 kHz indi-
vidually configured per channel. The measurement time can be chosen from a pool of 8 meas-
urement times ranging from 1.3 s to 163.48 s. Alternatively an undefined measurement time
can be set where multiple hours of measurement data can be collected depending on the
available free disk space and the measurement configuration.
Synchronously up to 2 speed channels with a constant sampling rate of 6.4 kHz can be meas-
ured.
The software Peakanalyzer Manager provides the possibility to configure and manage the
PeakStore. It also provides an interface to control the measurement and input additional side
information in a measurement protocol. Additionally, online spectra can be calculated during
the measurement.
1.1.
CE-Labeling
We declare herewith,
GfM Gesellschaft für Maschinendiagnose mbH
Köpenicker Straße 325, Haus 40, D-12555 Berlin,
that our product
PeakStore5
in all product variations
1
complies to the following relevant regulations:
2
,
3
1
This applies to complete measurement systems. For individual components the conformity has to be proven in
the target system. If applied correctly it is not expected to have affects according to device safety or EMV.
2
This conformity declaration applies only the systems supplied by GfM. Changes or extensions are in the respon-
sibility of the operator and thus has to ensure the conformity with the corresponding EG-Directive.
3
Harmonised Norms:
DIN EN 61326:2002 (Produktnorm): Es werden die Störfestigkeits-Prüfanforderungen an Betriebsmittel, die zum
Gebrauch in industriellen Bereichen vorgesehen sind (gemäß EN 61326/A1 Tabelle A1 und Tabelle 3) sowie die
Störaussendungsgrenzwerte (gemäß EN61326/Tab.3) eingehalten.
DIN EN 61000-4-2:2001, -4-3:2001, -4-4:2002, -4-5:2001, -4-6:2001
DIN EN 61010-1:2002
DIN EN 55011:2000 (Funkstörfeldstärke Klasse A)