PeakStore5 Manual Version 3.0 

6. 

Software Usage 

The software Peakanalyzer Manager serves as interface to manage the data collection sys-
tems  PeakStore504,  PeakStore508  and  PeakStore512.  At  first  the  license  of  the  according 
PeakStore5 must be added to the Peakanalyzer Manager. The license manager is available 
via the menu item Extras 

→ Licenses. Via the button Add License the license that is delivered 

with the PeakStore5 can be added.  

The license can be extended by a balancing module, allowing to balance machines with up to 
two balancing layers. If this option is activated, it is also shown in the license manager in the 
Peakanalyzer Manager.  

The coupling of the PeakStore5 to the tablet is done via the expansion port in the standard 
delivery. This expansion port is activated automatically during the boot process of the tablet 
but can also be activated manually via the tablet module configuration.  

If the PeakStore5 is used via the separate USB connection in combination with a normal PC, 
the  USB  connection  must  be  first  established  before  the  PeakStore5  is  added  to  the 
Peakanalyzer Manager.  

Afterwards  the  PeakStore5  can  be  added  to  the  Peakanalyzer  Manager  via  the  menu  item 
Diagnosis system 

→ New → PeakStore. The button Search lists all supported measurement 

devices that are connected via USB.  

The  PeakStore5  is than listed  in  the  device  browser  of  the  Peakanalyzer  Manager  like  any 
other system. After the selection of the PeakStore5 in the device browser, a query of the chan-
nel properties is done and the configuration site is opened.  

Via the menu Extras 

→ Options → PeakStore → General the directory where all measurement 

data should be saved can be chosen. Additionally, a shortcut to the desktop or in the Windows 
startup can be created. This will start the Peakanalyzer Manager and automatically select the 
desired PeakStore5 and will start the settling process with active configuration.  

6.1. 

Configuration 

In the configuration view, there is a menu bar in the bottom with some general functionalities. 
The button New will create a new configuration. The button Activate will choose the currently 
opened configuration as measurement configuration. This active configuration will also be au-
tomatically loaded if the PeakStore5 is selected in the device browser. The button Save will 
save all changes of the currently opened configuration if there are no validation errors. If invalid 
configuration parameters are set, a window will be opened with description of the invalid pa-
rameter. The button delete will delete the selected configuration. This is only possible if there 
are no measurement data existent for this configuration. If applicable, these have to be deleted 
in the data management first.  

Содержание PeakStore5

Страница 1: ...www maschinendiagnose de PeakStore5 Manual Version 3 0...

Страница 2: ......

Страница 3: ...25 12555 Berlin Phone 49 30 65 76 25 65 Fax 49 30 65 76 25 64 Dipl Ing Axel Haubold Sales office West Am Kaempen 22 44227 Dortmund Phone 49 231 77 64 723 Fax 49 231 77 64 724 Dipl Ing Kai Uchtmann Ban...

Страница 4: ...1 Sensor Inputs 9 4 2 Speed Inputs 10 4 3 External sensor supply voltage and protected earth connection 10 4 4 USB Connection 11 4 5 USB connection expansion port 11 4 6 Connections at the Tablet 12...

Страница 5: ...PeakStore5 Manual Version 3 0 3 6 3 2 Origin Measurement 27 6 3 3 Test Measurement 27 6 3 4 Validation Measurement 27 6 3 5 Finalization 28 6 4 Data Management 29 7 Remote Control 32...

Страница 6: ...an interface to control the measurement and input additional side information in a measurement protocol Additionally online spectra can be calculated during the measurement 1 1 CE Labeling We declare...

Страница 7: ...ed to the PeakStore5 must be shielded Please consider that all signal inputs and outputs except the power connection is shielded Only in this case a high interference immunity and a reduced interferen...

Страница 8: ...the device 1 8 Transportation Transport the PeakStore5 in the original package or in a package that ensures an adequate protection Transport damages are excluded from guarantee claims 1 9 Before Inst...

Страница 9: ...n Before exchanging the accumulator it is required to turn of the tablet The accumulator can be accessed on the back of the tablet Beforehand the measurement system must be unplugged This can be done...

Страница 10: ...24 V Cable for protected earth connection Replacement accumulator 3 Control Elements Picture 2 Function keys front bottom Source Winmate M101 Manual The measurement can be started via the function key...

Страница 11: ...al power supply has to be connected to the DC IN connector on the right side see 4 3 While connecting the sensors the correct assignment of the numbered sensors to the inputs has to be considered The...

Страница 12: ...of the PeakStore5 or via the remote control The display can variate slightly If there are heavy variations the orientation of the speed sensor should be checked and the measurement should be repeated...

Страница 13: ...from GfM It has to be ensured that not both the expansion port and the separate USB connection are used in parallel Picture 8 Pin assignment USB connection 4 5 USB connection expansion port The connec...

Страница 14: ...measurements 5 Standby Mode To achieve a long battery runtime it is wise to switch of the measurement module if it is not needed There are two different modes available to reduce the power consumptio...

Страница 15: ...life time of the tablet is increased considerably by this deactivation However the reactivation time for the measurement module is much longer than the other variant To deac tivate the expansion port...

Страница 16: ...ted measurement devices that are connected via USB The PeakStore5 is than listed in the device browser of the Peakanalyzer Manager like any other system After the selection of the PeakStore5 in the de...

Страница 17: ...lculation of online order and frequency spectra is desired during the measurement If the online calculation is not activated the battery life time is increased Show IP Address for App If this option i...

Страница 18: ...the impulses per rotation that are measured by the sensor must be adjusted to the correct value The trigger threshold of the speed channel can be adopted if necessary The speed sensors that are deliv...

Страница 19: ...or For accelerometers with IEPE interface this value is usually 0 For analogues displacement sensors the offset of the displacement sensor must be entered Unit the physical unit of the sensor Measurem...

Страница 20: ...onfigured for this sensor The following r m s values of vibration velocity or vibration acceleration can be configured for each sensor DIN ISO 10816 3 with bandwidth 10 Hz 1 kHz DIN ISO 10816 21 with...

Страница 21: ...owing templates are availa ble 2 Spur gear stage Planetary stage 3 Spur gear stage 4 Spur gear stage Planetary stage A60 B130 Blade bearing Motor Planetary stage 2 Spur gear stage Motor Motor Gear box...

Страница 22: ...s also increases the power consumption of the PeakStore5 If the monitoring mode is not needed it is recommended to deactivate the PeakStore5 to increase the battery life time A measurement can be star...

Страница 23: ...ors and other sensors The current measurement value is shown Speed The current speed value Additionally the variation of the speed in percent during the measurement By pressing on the sensor the time...

Страница 24: ...If an already placed sensor is clicked again a popup will open to change the measurement position and direction If the same sensor arrangement is used in multiple measurements it is recommended to sav...

Страница 25: ...sors also in parallel can be viewed The signals that should be shown can be changed by the sensor listing on the right side The curve window is frequently updated to show the latest data during the me...

Страница 26: ...visualization of the characteristic values can be found on the characteristic value site The characteristic value is recalculated continuously during the measurement for each sensor The graphical vie...

Страница 27: ...fterwards up to 5 validation measurements will follow where balancing masses are applied to the layers to reduce the initial imbalance Generally each measurement consists of a preparation phase the ac...

Страница 28: ...e vibration sensor must be placed on each layer It is necessary to take into account the correct layer for each vibration sensor as parametrized in the configuration see also 6 1 3 Picture 23 Paramete...

Страница 29: ...re set up It will be first waited until the speed is in the valid range If this is the case the actual measure ment starts automatically and also stops automatically if the required number of revoluti...

Страница 30: ...ss The additionally cal culated masses must be added to the existing masses from the last validation measurement This step can be repeated up to 5 times 6 3 5 Finalization If the target quality level...

Страница 31: ...uration View The measurement data is grouped according to the used configura tion Order View The measurement data is grouped according to the order number that was entered in the measurement protocol...

Страница 32: ...here the desired target directory can be chosen Thus a con nected USB Stick can be chosen to export the data The selected measurement data will be compressed to a single file for all measurements or a...

Страница 33: ...ng and calculation Picture 30 Diagnostic Tool On the left side of the Diagnostic Tool the measurement data that were selected in the PeakStore5 data management are shown Each measurement can be prepro...

Страница 34: ...onnection must be established between Smartphone and PeakStore5 to get the PeakStore Remote App work ing It is recommended to use an additional WLAN route as access point between these two where both...

Страница 35: ...and maximal values are only updated during a measurement If the measurement is currently stopped only the current value is updated Via the button Start on the right bottom a measurement can be started...

Страница 36: ......

Страница 37: ...Q0188e GfM Gesellschaft f r Maschinendiagnose mbH Koepenicker Strasse 325 12555 Berlin Deutschland Germany Phone Berlin 49 30 65 76 25 65 Phone Dortmund 49 231 77 64 723 www maschinendiagnose de...

Отзывы: