Signet 5500 Flow Monitor
9.3 OPTIONS
Menu
Choose:
1
Change:
2
Save:
3
quick
press
To return to
VIEW menu:
quick
press
To restore original value:
Display contrast
Low -- -- -- -- -- High
4 mA adjust
Loop output
Loop output
20 mA adjust
Flow display
Decimal
Press
will appear during code entry.
keys in sequence to enter menu,
Total display
Total decimal
Display averaging
4s
8s
0s
Loop output
4 mA setpoint
20 mA setpoint
Loop output
J↓
A↑
ENTER
"SAVING"
briefly
displays
Totalalizer reset
VIEW menu=
VIEW menu=
A.
B.
C.
D.
E.
F.
G.
H.
I.
J.
Auxiliary out logic
Active Low or High
Menu Settings A - J:
(Factory default displays shown)
A. Selects LCD display contrast: 5 levels
B. Selects
fl
ow display decimal:
*****. to ** . ***
C. Selects totalizer display decimal:
********. to ****** . **
D. Selects LCD display averaging:
Off = 0 seconds, Low= 4 seconds, High= 8
seconds (also affects 4 to 20 mA output)
E. Sets 4 mA output setpoint. 4 mA and 20 mA
setpoints are reversible.
F. Sets 20 mA output setpoint. 20 mA and 4 mA
setpoints are reversible.
G. Sets 4 mA current output: 3.0 to 5.0 mA
(overrides 4.00 mA factory calibration)
H. Sets 20 mA current output: 19 to 21 mA
(overrides 20.00 mA factory calibration)
I. Selects totalizer reset options: Lock "Off"
disables the VIEW menu access code
feature; Lock "On" enables the feature (sec.
9.1 A). When enabled, the operator must
enter an access code for totalizer reset
privileges. The totalizer will be reset in either
case if a contact closure is detected at rear
reset terminals (sec. 5).
J. Selects Auxiliary output logic:
• Active Low logic (factory default): Output
changes from high to low when active, see
diagram below.
• Active high logic: Output changes from low
to high when active, see diagram below.
Inactive (Off)
Active Low Logic (factory default)
Active (On)
Inactive (Off)
Active High Logic
Active (On)