Signet 5500 Flow Monitor

Relay mode 
and setpoint 
displays repeat 
for relay 2 
setup

CALIBRATE Menu Setup Parameters (sec. 9.2)

_ = blank 
display digit 
option

_ = blank 
display digit 
option

Menu Parameters

Display Description

Range

Factory Default

A.

Flow unit = digits 1-3
Flow timebase = digit 4

Aa - Zz, 0 - 9, _;

Mm, Ss, Hh, Dd

_GPM

B.

Min 

 max range

00.000 - 99999.

00.000 - 100.00

GPM

C.

Flow K-Factor

0.0001 - 99999.

60.000

D.

Total units

Aa - Zz, 0 - 9, _

(8 digits max)

_Gallons

E.

Total K-Factor

0.0001 - 99999.

60.000

F.

•  Relay 1 mode
•  Relay 1 setpoint

Low or High, 0.0000 - 99999.

Pulse, 0.0001 - 99999.

Low

10.000 GPM

G.

•  Relay 1 Hysteresis
• Hysteresis volume

0.0000 - 99999.

2.0000

GPM

H.

•  Relay 1 mode
•  Relay 1 pulse volume

Pulse

0.0001 - 99999.

10.000

Gallons

I.

Relay 1 pulse width

0.10 - 999. seconds

10.0

seconds

J.

•  AUX output mode
•  AUX output setpoint

Low or High, 0.0000 - 99999.

Low

1.0000

K.

AUX output hysteresis 
volume

0.0000 - 99999.

0.0000

L.

•  AUX output mode
•  AUX output setpoint

Pulse, 0.0001 - 99999.

1.0000

Gallons

M.

AUX output pulse width

0.10 - 9999. seconds

0.10 seconds

N.

Last calibration date

00-00-00 to 39-39-99

01-01-98

Содержание Signet 5500

Страница 1: ...Double Insulated DC or AC power Contents 1 Power Connections 2 Compatible Sensor Wiring 3 Sensor Pulse Output Connections 4 Auxiliary Pulse Output Connections 5 Totalizer Reset Connections 6 4 20 mA...

Страница 2: ...t PNP source input 5500 Terminals External power supply 28 VDC max Other instrument Gnd Input NPN sink input Gnd AUX output N 5500 Terminals Signal output Signal ground 3 28 VDC External power supply...

Страница 3: ...wer is applied after the relay 2 setpoint is reached 6 4 to 20 mA Current Output Connections sec 8 4 4 20 mA 5500 Terminals 4 20 mA Fuse 1 8A Technical Notes 1 8 A fuse recommended customer supplied 4...

Страница 4: ...nsor also displayed on totalizer To enable the totalizer function the operator must set the relay or auxiliary pulse volume setpoint to 1 0000 sec 9 2 H 9 2 L Totalizer function ideal for external cou...

Страница 5: ...djustments ENTER OPTIONS sec 9 3 CALIBRATE sec 9 2 VIEW sec 9 1 2 seconds 5 seconds Press hold ENTER key for menu access Menu Displays A H Factory default displays shown A Flowrate and resettable tota...

Страница 6: ...disable hysteresis feature L Sets auxiliary output pulse volume setpoint 0 0001 99999 sec 8 2 Set to 1 0000 factory default to enable totalizer function sec 8 2 M Sets auxiliary output pulse width 0...

Страница 7: ...nds High 8 seconds also affects 4 to 20 mA output E Sets 4 mA output setpoint 4 mA and 20 mA setpoints are reversible F Sets 20 mA output setpoint 20 mA and 4 mA setpoints are reversible G Sets 4 mA c...

Страница 8: ...ilicone rubber Weight Approximately 500 g 18 oz Display Type Microprocessor controlled air core meter movement and backlit alphanumeric 2 x 16 LCD Update rate Flow 1s totalizer 200 ms Contrast User se...

Страница 9: ...96 mm 3 8 in RELAY 2 RELAY 1 ENTER 4 0 1 2 3 5 6 7 8 9 10 GPM Signet Monitor 92 x 92 mm 3 62 x 3 62 in Panel cutout Menu Parameters Display Description Range Factory Default A Flow rate and units Rese...

Страница 10: ...max _Gallons E Total K Factor 0 0001 99999 60 000 F Relay 1 mode Relay 1 setpoint Low or High 0 0000 99999 Pulse 0 0001 99999 Low 10 000 GPM G Relay 1 Hysteresis Hysteresis volume 0 0000 99999 2 0000...

Страница 11: ...l to D Display averaging Off 0 sec Low 4 sec High 8 sec Off E 4 mA setpoint Unlimited 0 0000 GPM F 20 mA setpoint Unlimited 100 00 GPM G 4 mA adjust 3 0 to 5 0 mA 4 00 mA H 20 mA adjust 19 to 21 mA 20...

Страница 12: ...mall Reduce auxiliary output pulse width setting sec 9 2 M or increase pulse volume setting sec 9 2 L Must be between 0 0001 99999 Relay 1 Relay 2 or Auxiliary output AUX pulse volume setpoint must be...

Отзывы: