10
Signet 5500 Flow Monitor
Relay mode
and setpoint
displays repeat
for relay 2
setup
CALIBRATE Menu Setup Parameters (sec. 9.2)
_ = blank
display digit
option
_ = blank
display digit
option
Menu Parameters
Display Description
Range
Factory Default
A.
Flow unit = digits 1-3
Flow timebase = digit 4
Aa - Zz, 0 - 9, _;
Mm, Ss, Hh, Dd
_GPM
B.
Min
→
max range
00.000 - 99999.
00.000 - 100.00
GPM
C.
Flow K-Factor
0.0001 - 99999.
60.000
D.
Total units
Aa - Zz, 0 - 9, _
(8 digits max)
_Gallons
E.
Total K-Factor
0.0001 - 99999.
60.000
F.
• Relay 1 mode
• Relay 1 setpoint
Low or High, 0.0000 - 99999.
Pulse, 0.0001 - 99999.
Low
10.000 GPM
G.
• Relay 1 Hysteresis
• Hysteresis volume
0.0000 - 99999.
2.0000
GPM
H.
• Relay 1 mode
• Relay 1 pulse volume
Pulse
0.0001 - 99999.
10.000
Gallons
I.
Relay 1 pulse width
0.10 - 999. seconds
10.0
seconds
J.
• AUX output mode
• AUX output setpoint
Low or High, 0.0000 - 99999.
Low
1.0000
K.
AUX output hysteresis
volume
0.0000 - 99999.
0.0000
L.
• AUX output mode
• AUX output setpoint
Pulse, 0.0001 - 99999.
1.0000
Gallons
M.
AUX output pulse width
0.10 - 9999. seconds
0.10 seconds
N.
Last calibration date
00-00-00 to 39-39-99
01-01-98