40
Vapodest 45s
9 Trouble shooting
Error message
Reason for error
Actions to be taken
Steam generator
Low water
Not enough water in the
steam generator, water
inlet might be broken
Check tubing connection from water tank to Vapo-
dest
Should the problem reoccur, call customer service.
9.2. Breakdown of operation
Steam generator
Sensor Error
Water level of steam ge-
nerator is not measured
any longer.
Turn instrument off, and then turn it on again. Should
the problem reoccur, call customer service.
Error message
Reason for error
Actions to be taken
Excess current switch
cuts off
Excessive current con-
sumption
Turns switch on again.
Should the problem reoccur, call customer service.
Steam generator
Excess pressure
Water level in steam ge-
nerator too low.
Heating starts too early.
Turn off instrument for ca. 15 min and turn on again.
Should the problem not be solved, call customer
service.