37
Vapodest 45s
8. The display shows the individual program steps.
You can interrupt the program at any time, during
the distillation in process. Please make sure to
read chapter
"8.4. Interrupt distillation"
.
9
. At the end of the analysis, the result of the titra-
tion is shown. Sample and receiver are sucked
off automatically. press the fast clamping device
downwards, if you want to take out the digestion
tube.
Careful ! At the end of the distillation, the digestion
tube is very hot. Thus make sure to wear protective
gloves and protective goggles!
10.
Would you like to continue with the program and
run additional.
Analysis, then please insert a new digestion tube
and press the push button. The receiver will be
filled with boric acid again.
Should you not wish any additional analysis, then
turn the push button in any direction and select
"Back".
8 Execution of method
M01 Adding H2O
100ml
M01 Adding NaOH
080ml
M01 Reaction
00min 00s
M01 Distillation
03min 30s
M01 Titration
EP pH: X,XXX
M01 Suction time
P:20s V:10s
M01 Adding H3BO3
080 ml
M01 Ammonium
X.XXXml pH X.XX
press
M01 Ammonium
Exit
turn
M01 pH X.XX
Stop Continue
press
M01 pH: X.XX
pH-measurement