7
Vapodest 45s
3.1. Warranty conditions
Vapodest 45s is designed and fabricated following the high quality guidelines of
DIN EN ISO 9001.
On the basis of the C. Gerhardt conditions of warranty, our products are guaranteed
for 12 months as long as the equipment is used according to the instructions in
this manual. Also fill in the "Report for the installation and initial operation" and
return it to C. Gerhardt.
Please note that the following parts are excluded from warranty.
- Digestion tube, Order-no.: 1002178
- Vitone cone, Order-no.: 12-0351
3.2. Technical data
Nominal voltage:
230 V AC, 50 Hz or 115 V 60 Hz or 230 V 60 Hz
Nominal wattage:
1600 W
Cooling water consumption:
about 5 liters per distillation minute
Cooling water pressure:
min. 0,5 bar, max. 6 bar
Pump capacity (read chapter 6.3.2. Calibration dosing pumps)
- Diaphragm pump about 15 ml/s with water
about 10 ml/s with NaOH
- Peristaltic pump about 10 ml/s with water
Storage tanks:
any size, recommended:
Set of storage tanks KAN 40
Display:
LED with illumination
Programs:
20
Dimensions:
440 x 690 x 340 mm (W x H x D)
Weight:
29 kg
Interfaces:
- Vapodest
2 x RS 485 interface: Connection for GerBus®
1 x RS 232 interface: Connection for titrator
1 x 5-pin connector for level sensor storage tanks
3 Technial description