Instructions for Use
Ergostik Blueflow
Serial no.: xx|8|201|yyy and 2203xxxxx
Version: 6 | Release date: 02 August 2022
Page 79
Error
Rectification
No flow signal
Check that the Blueflow is correctly positioned in
the Blueclip
Check the tube connection between Blueclip and
Ergostik Blueflow
Drift on flow signal
Check that the Blueflow is correctly positioned in
the Blueclip
Check the tube connection between Blueclip and
Ergostik Blueflow
Check for moisture in the tube / connection
piece of the Blueflow
Perform volume calibration
No gas signal
Check that the Blueflow is correctly positioned in
the Blueclip
Check the tube connection between Blueclip and
Ergostik Blueflow
Check for moisture in the tube / connection
piece of the Blueflow
Perform gas calibration
Drift on gas signal
Note the warm-up time of the Ergostik Blueflow
(15 minutes)
Check for moisture in the tube / connection
piece of the Blueflow
Perform gas calibration
Software error message
"Unit warm-up not
completed.
Note the warm-up time of the Ergostik Blueflow
(15 minutes)
Software error message
"Flow out of measuring
range".
Check that the Blueflow is correctly positioned in
the Blueclip
Check tube connection between Blueclip and
Ergostik Blueflow