Version: 6 | Release date: 02 August 2022
7.2
Establishing Operational Readiness
To make the Ergostik Blueflow ready for operation, please read
the instructions:
•
Chap. 7.5ff "Calibrating Ergostik Blueflow“
7.3
Switching Ergostik Blueflow On / Off
Possible physical injury.
Reason: misinterpretation of obvious measurement
errors (drift). Therefore:
•
Let the Ergostik Blueflow warm up for
15 minutes! Measurements may only be
performed after this time!
The Ergostik Blueflow can be switched off by disconnecting the
power supply. When using a cart, this is done via the On / Off
switch on the cart. Otherwise, the On / Off switch on the back of
the Ergostik Blueflow must be activated.
To extend the life of the gas sensors, the Ergostik Blueflow
should be switched off if there are long intervals between
measurements.
After switching on the Ergostik Blueflow needs a warm-up time
of 15 minutes. Only after this time is the Ergostik Blueflow ready
for operation.
Detailed information on how to establish operational readiness
you will find here:
•
In this IFU – chap. 7.2 "Establishing operational
Readiness".
If the calibration as described in chap. 7.5 "Calibrating Ergostik
Blueflow" has not been carried out already, this must be done
later, otherwise no functional readiness is possible.