Instructions for Use
Ergostik Blueflow
Serial no.: xx|8|201|yyy and 2203xxxxx
Version: 6 | Release date: 02 August 2022
Page 85
12.2
Installation and Operating Conditions
The following conditions supplement chap. 2.2 "Intended Use"
and must be observed to maintain the properties of the Ergostik
Blueflow guaranteed by the manufacturer and for the safety of
the patient and operator.
Storage / Transport:
min.
max.
Temperature:
-10 °C
+50 °C
Relative air humidity:
10 %
95 %
non-condensing
Atmospheric pressure:
700 hPa
1100 hPa
Operation:
min.
max.
Environmental temperature:
+10 °C
+35 °C
Avoid extreme
fluctuations in
temperature!
Relative air humidity:
20 %
95 %
non-condensing
(at least 30 % for synthetic flooring)
Atmospheric pressure:
700 hPa
1100 hPa
Setup:
mobile
Space requirement:
(complete with medical cart)
(L) 900 mm x (W) 900 mm x (H) 1900 mm
Required floor load capacity: 150 kg/m²
Flooring:
stable; even; conductive
preferably no synthetic material
Environment / room:
Closed, clinical area
air-conditioned
not an explosive or flammable environment!
not near an MRI!
well ventilated
Operating Mode:
Continuous operation