Instructions for Use
Ergostik Blueflow
Serial no.: xx|8|201|yyy and 2203xxxxx
Page 80
Version: 6 | Release date: 02 August 2022
Error
Rectification
Check for moisture in the tube / connection
piece of the Blueflow
Software error message
"O
2
out of measuring
range“
Check tube connection between Blueclip and
Ergostik Blueflow
Check for moisture in the tube / connection
piece of the Blueflow
Check gas outlet for blockage
Perform gas calibration
Check the set type of the O
2
Sensor
Software error message
"CO
2
out of measuring
range
Check tube connection between Blueclip and
Ergostik Blueflow
Check for moisture in the tube / connection
piece of the Blueflow
Check gas outlet for blockage
Perform gas calibration
obvious measuring error Check that the Blueflow is correctly positioned in
the Blueclip
Check tube connection between Blueclip and
Ergostik Blueflow
Check for moisture in the tube / connection
piece of the Blueflow
Check the tightness of the Mask / Mouthpiece
Check last calibration date and recalibrate if
necessary (see chap. 7.5)