Instructions for Use
Ergostik Blueflow
Serial no.: xx|8|201|yyy and 2203xxxxx
Page 42
Version: 6 | Release date: 02 August 2022
Reason: Disregard of a contraindication.
Misdiagnosis due to measurement errors. For this:
Observe general medical principles! Therefore:
•
Inform yourself and observe the respective
contraindications before each test!
•
When carrying out tests, observe the content of
the applicable guidelines and recommendations
(e.g. ATS / ERS Guidelines)!
•
Do not use single use products more than once!
Reason: Side-effects (i.e. cardiac arrythmia) triggered
by exercise. Therefore:
•
Perform exercise tests only in presence of a
physician!
•
Ensure availability of resuscitation equipment
(defibrillator, medications) and trained emergency
personnel!
•
Monitor cardiac function with appropriate
methods (e.g. ECG) during exercise test!
3.4
Electromagnetic Compatibility (EMC)
Possible danger to life.
Reason: Misdiagnosis due to measurement error
caused by a system failure due to uncontrollable
electromagnetic fields of inadmissible transmitting
devices. Therefore:
•
While using Ergostik Blueflow, do not use any
transmitting devices (e.g. mobile phones, portable
phones, power lines …) that exceed the immunity
levels as specified in the EMC guidelines!
•
Ask your authorised specialist retail partner!