Guidance and manufacturer´s declaration – electromagnetic emissions
The device is intended for use in the electromagnetic environments listed below, and
should only beused in such environments:
Emissions test
Compliance
Electromagnetic environment - guidance
RF emissions CISPR 11
Group 1
RF energy is used only to maintain device’s
operation. Therefore, its RF emissions are so
low that it’s not likely to cause any interfer-
ence in nearby electronic equipment.
RF emissions CISPR 11
Class B
The device is suitable for use in all establish-
ments, including domesticestablishments,
and those directly connected to the public
low-voltage power supply networkthat sup-
plies buildings used for domestic purposes.
Harmonic emissions
IEC 61000-3-2
Class A
Voltage
fl
uctuations /
fl
icker
emissions IEC 61000-3-3
complies
Guidance and manufacturer´s declaration – electromagnetic emissions
The device is intended for use in the electromagnetic environments listed below, and
should only beused in such environments:
Immunity test
IEC 60601 –
test level
Compliance
level
Electromagnetic environment -
guidance
Electrostatic
discharge (ESD)
IEC 61000-4-2
± 6 kV contact
± 8 kV air
± 6 kV contact
± 8 kV air
Floors should be wood, concrete
or ceramic tile. If
fl
oors are covered
with synthetic material, the relative
humidity should be at least 30 %.
Power frequency
(50/60 Hz)
magnetic
fi
eld
IEC 61000-4-8
3 A/m
3 A/m
Power frequency magnetic
fi
elds
should be at levels characteristic of
a typical location in a typicalcommer-
cial or hospital environment.
Recommended separation distances between portable and mobile RF communication
The device is intended for use in an electromagnetic environment where radiated RF
disturbancesare under control. User can help prevent electromagnetic interference by
keeping the device at a minimum distance from portable and mobile RF communications
equipment (transmitters). Below table details the maximum output power of transmitter:
Rated maximum
output power
of transmitters
in Watt
Seperation distance according to frequency of transmitter / m
150 kHz to 80 MHz
d = 1.2
√
P
80 MHz to 800 MHz
d = 1.2
√
P
800 MHz to 2.5 GHz
d = 2.3
√
P
0.01
0.1
1
10
100
0.12
0.38
1.2
3.8
12
0.12
0.38
1.2
3.8
12
0.23
0.73
2.3
7.3
23
For transmitters rated at a maximum output power not listed above, the recommended
separation distance d in metres (m) can be estimated using the equation applicable to the
frequency of the transmitter, where P is the maximum output power rating of the transmit-
ter in watts (W) according to the transmitter manufacturer.
NOTE 1:
At 80 MHz and 800 MHz, the separation distance for the higher fre-
quency range applies.
NOTE 2:
These guidelines may not apply in all situations. Electromagnetic
propagation is affected by absorption and re
fl
ection from structures,
objects and people.
Содержание desktop 2.0
Страница 2: ......
Страница 3: ...DE Digitaler Blutdruckmonitor f r den Oberarm Gebrauchsanweisung GT 6630...
Страница 28: ......
Страница 29: ...EN Digital Blood Pressure Monitor for the Upper Arm INSTRUCTION FOR USE GT 6630...
Страница 54: ......
Страница 55: ...FR Tensiom tre lectronique au bras MODE D EMPLOI GT 6630...
Страница 80: ......
Страница 81: ...IT Misurapressione digitale da braccio ISTRUZIONI PER L USO GT 6630...
Страница 106: ......
Страница 107: ...ES Monitor digital de tensi n arterial de brazo INSTRUCCIONES DE USO GT 6630...
Страница 132: ......
Страница 133: ...PT Monitor digital de tens o arterial de bra o INSTRU ES DE UTILIZA O GT 6630...
Страница 158: ......
Страница 159: ...NL Digitale bloeddrukmonitor voor de bovenarm BEDIENINGSHANDLEIDING GT 6630...
Страница 184: ......
Страница 185: ...AR GT 6630...
Страница 186: ...AR 186 LCD 186 187 189 190 192 193 193 194 194 195 196 197 198 198 200 200 201 desktop 2 0 202 204 206 207 208 209...
Страница 187: ...AR 187...
Страница 188: ...AR 188 DIN EN ISO 81060 2 Power...
Страница 189: ...AR 189...
Страница 191: ...AR 191 mmHg...
Страница 192: ...AR 192 32 22 LCD USB...
Страница 193: ...AR 193 Geratherm desktop Double Check Technology fuzzy logic USB Geratherm desktop WHO...
Страница 194: ...AR 194 AA 4...
Страница 195: ...AR 195 1 2 2 5 3...
Страница 196: ...AR 196 Geratherm AC...
Страница 197: ...AR 197 MEMORY Power MEMORY...
Страница 198: ...AR 198 MEMORY START MEMORY START month fuzzy logic Double Check Technology 1 Double Check Technology fuzzy logic...
Страница 199: ...AR 199 START WHO START WHO...
Страница 200: ...AR 200 MEMORY MEMORY 1 MEMORY 2 ME MORY START 3 dEL START 4...
Страница 201: ...AR 201 1 MEMORY 2 START 3 dEL MEMORY 4 dEL ALL START 5 desktop 2 0 http geratherm de diagnostik downloadbereich...
Страница 202: ...AR 202...
Страница 203: ...AR 203 Power...
Страница 204: ...AR 204...
Страница 207: ...AR 207...
Страница 208: ...AR 208 BF mm yyyy SN 25 C 55 C 90 15...
Страница 209: ...AR 209 EMV EN60601 1 2 EMV Geratherm Medical EN 60601 1 2 2 3 3...
Страница 212: ...AR 212 IEC 60601 IEC 61000 4 6 3 150 80 3 d 1 2 P IEC 61000 4 3 3 80 2 5 3 800 d 1 2 P 80 d 2 3 P 800 2 5 P W d m a b...
Страница 213: ...AR 213 a b Geratherm Medical AG Fahrenheitstr 1 Geschwenda 98716 Deutschland 800 80 1 2...