32
3.22 NAddd
Sets the current network address to any address between 1 and 256.
*NA10
Network address: 10
*
When network recognition is enabled, this number (preceded by the # character) must be entered
for the respective datalogger to respond. The following example illustrates communication with
two different dataloggers on the RS-485 network.
<ENTER>
<ENTER>
#1<ENTER>
Network address: 1
*NA
Network address: 1
*E
<ENTER>
<ENTER>
#10<ENTER>
Network address: 10
*NA
Network address: 10
*E
NOTE: If connected directly to the datalogger via USB and networking is enabled, the
datalogger will respond with the * prompt only. The network address may not be changed while
networked. Direct connect to the datalogger via USB in order to change the network address.
3.23 ND
N
etwork
D
isable the datalogger. Disables networking of two or more LC-2x16 dataloggers.
*ND
Network recognition disabled.
3.24 NE
N
etwork
E
nable the datalogger. Enables networking of two or more LC-2x16 dataloggers.
*NE
Network recognition enabled.
Note: If the LC-2x16 is connected via the USB port, connection to a network enabled datalogger
can be made directly without the need to enter the correct datalogger address. This can be helpful
if the network address is unknown and the datalogger is network enabled.