34
3.29 S
Display the datalogger
S
tatus.
*S
MS:1004 OP:1005 UP:1004
16 Channel Multiplexer Selected.
Scan interval: 60 second(s).
Logging stopped.
Log intervals disabled.
Monitor mode enabled.
*
Line Description Manual
Sections
1
Status of memory pointers
3.7, 3.26
2 Multiplexer
Type 3.18
3
Scan interval setting
3.30
4
Start/Stop status
3.32, 3.34
5
Stop time (optional)
3.32
6
Log interval status
3.13, 3.14
7
Monitor mode status
3.15, 3.16, 3.17
Table 6 - S Command Information
3.30 SCnnnnn
Enter the
SC
an interval, in seconds. Range of entry is 3 to 86400. Only whole numbers are
accepted. Typing
SC
with no value returns the current setting only
*SC
Scan interval: 60 second(s).
*SC300
Scan interval: 300 second(s).
*
3.31 SS
Display the
S
ystem
S
tatus of the datalogger.
*SS
Trap count: 0
Network address: 1
Network recognition disabled.
Time format is hh,mm.
Date format is month,day.
*