39
3.38 Tnn/t
Enter the channel’s (nn) thermistor type (t). This command allows a specific thermistor to be
assigned to each individual channel (generally the external thermistor that is incorporated into
the VW gage). Entering T1/0 sets the external thermistor type of Channel 1 to the standard
3K
Ω
@25ºC NTC (default). For example, entering T2/1 sets the external thermistor type of
Channel 2 to the high temperature BR55KA822J 8.22K
Ω
@25ºC NTC thermistor. Entering T3/2
sets the external thermistor type of Channel 3 to the high temperature 103JL1A 10K
Ω
@25ºC
NTC thermistor.
T1/0:
CH 1: ENABLED
GT: 1 ZR: 0.00000 GF: 1.00000 GO: 0.00000 TH: 0
T2/1:
CH 2: ENABLED
GT: 1 ZR: 0.00000 GF: 1.00000 GO: 0.00000 TH: 1
T3/2:
CH 3: ENABLED
GT: 1 ZR: 0.00000 GF: 1.00000 GO: 0.00000 TH: 2
3.39 TR
Display the current
TR
ap Count. The trap counter is a register that keeps track of the number of
times that the internal processor has detected a communications error. This is a useful register to
check if communication problems are suspected.
3.40 TR0
Reset the
TR
ap count register to
0
.
3.41 VL
Display the Lithium Coin Cell Voltage. The internal 3V lithium coin cell is used to supply power
to the real-time clock circuit. The 3V lithium coin cell life is rated at 10 years minimum.
*VL
Lithium Cell Voltage = 2.92V
*
3.42 V3
Display the 3V D-cell battery pack voltage. Replace the batteries when this voltage is less than
1.8V
*V3
3V Battery Voltage = 2.93V
*