Table 6 - S Command Information
3.30 “SCnnnnn” - Set Scan Interval
Enter the scan interval, in seconds. Range of entry is 3 to 86400 and is dependent on the number
and type of transducers connected. Only whole numbers are accepted. Typing
SC
with no value
returns the current setting only
*SC
Scan interval: 60 second(s).
*SC300
Scan interval: 300 second(s).
*
3.31 “SS” - Display System Status
Displays the System Status of the datalogger.
*SS
Trap count: 0
Network address: 1
Network recognition disabled.
Time format is hh,mm.
Date format is month,day.
*
Line
Description
1
Trap Count (Communication errors counter)
2
Current network address
3
Current network status
4
Current time format configuration.
5
Current date format configuration.
Table 7 - SS Command Information